Cook Incorporated · Nitinol Tipped Stone Extractor · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00827002362505), received between Jun 26, 2017 and Mar 12, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 5, 2026Open, Classified

    Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

    Z-1787-2026 · Root cause: Incorrect or no expiration date · FDA record ↗

Most reported problems

Device problems

  • Difficult to Open or Close7
  • Break2
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2026Material Separation

    IT WAS REPORTED THAT AN N-COMPASS NITINOL STONE EXTRACTOR BASKET WAS BROKEN. THE SURGEON WAS UNAWARE THAT THE BASKET HAD BROKEN AND WAS UNSURE WHEN THE BREAKAGE OCCURRED. AN X-RAY WAS PERFORMED ON THE PATIENT TO CONFIRM THAT NO PIECES OF THE BASKET REMAINED IN SITU. NO PORTION OF THE DEVICE REMAINED INSIDE OF THE PATIENT'S BODY. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579853 ↗
  • MalfunctionOct 27, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT, UPON OPENING THE PACKAGE PRIOR TO THE PROCEDURE, IT WAS FOUND THAT THE N-COMPASS NITINOL STONE EXTRACTOR COULD NOT BE OPENED. A NEW LIKE-DEVICE WAS OPENED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393536 ↗
  • MalfunctionOct 27, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT, UPON OPENING THE PACKAGE PRIOR TO THE PROCEDURE, IT WAS FOUND THAT THE N-COMPASS NITINOL STONE EXTRACTOR WAS UNABLE TO BE OPENED OR CLOSED. A NEW LIKE-DEVICE WAS OPENED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. A CUSTOMER PROVIDED PHOTOGRAPH SHOWS THE HANDLE HAS BEEN DISASSEMBLED. IT IS POSSIBLE THAT WHEN ATTEMPTING TO DISCONNECT THE HANDLE FROM THE HOOP, THE USER UNSCREWED THE HANDLE FROM THE MLLA (MALE LUER LOCK ADAPTER) INSTEAD OF UNSCREWING THE MLLA FROM THE RED HOOP CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393463 ↗
  • MalfunctionOct 27, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT A N-COMPASS NITINOL STONE EXTRACTOR WAS UNABLE TO BE OPENED OR CLOSED PRIOR TO AN UNKNOWN PROCEDURE FOR AN UNKNOWN PATIENT. UPON OPENING THE PACKAGE, IT WAS FOUND THAT THE PLASTIC TUBE HAD DETACHED, CAUSING THE INABILITY OF THE BASKET TO OPEN OR CLOSE. A NEW LIKE-DEVICE WAS OPENED TO COMPLETE THE PROCEDURE SUCCESSFULLY. A CUSTOMER PROVIDED PHOTOGRAPH SHOWS THE HANDLE HAS BEEN DISASSEMBLED. IT APPEARED LIKELY THE USER UNSCREWED THE HANDLE FROM THE MLLA ON THE DISTAL END OF THE HANDLE, INSTEAD OF UNSCREWING THE MLLA FROM THE RED CONNECTOR ON THE HOOP. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393320 ↗
  • MalfunctionAug 21, 2025Break; Break

    IT WAS REPORTED THAT UPON OPENING THE PACKAGE OF A N-COMPASS NITINOL STONE EXTRACTOR, THE "HOLDER OF THE SHEATH" WAS DISCOVERED TO BE BROKEN. THE DEVICE WAS NOT USED, AND ANOTHER DEVICE WAS OBTAINED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22860500 ↗
  • MalfunctionAug 21, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT UPON OPENING THE PACKAGE OF A N-COMPASS NITINOL STONE EXTRACTOR, THE BASKET FAILED TO RELEASE. ANOTHER DEVICE WAS OBTAINED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22860451 ↗
  • MalfunctionJul 1, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT UPON OPENING A N-COMPASS NITINOL STONE EXTRACTOR IT WAS DISCOVERED THE HANDLE WAS INOPERABLE. THE DEVICE DID NOT MAKE PATIENT CONTACT. A NEW LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22371873 ↗
  • MalfunctionJul 1, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT UPON OPENING A N-COMPASS NITINOL STONE EXTRACTOR IT WAS DISCOVERED THE BASKET DID NOT OPEN. THE DEVICE DID NOT MAKE PATIENT CONTACT. A NEW LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22370476 ↗
  • MalfunctionDec 16, 2019Difficult to Open or Close; Difficult to Open or Close

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 9475503 ↗
  • MalfunctionJun 26, 2017Break; Break

    INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT WAS SCHEDULED TO UNDERGO A CHOLEDOCHOTOMY USING A N-COMPASS NITINOL STONE EXTRACTOR. THE USER FOUND THAT THE OUTER SHEATH OF THE DEVICE NEARBY THE HANDLE END WAS BROKEN. THE EVENT WAS NOTED PRIOR TO USE DURING PROCEDURAL PREPARATION. PHOTOS SUPPLIED BY THE CUSTOMER SHOW A SEVERING OF THE CATHETER SHEATH JUST DISTAL TO THE FLARE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE ARE NO REPORTED ADVERSE PATIENT CONSEQUENCES. THE DEVICE IS AVAILABLE FOR EVALUATION, HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURE AS OF THE DATE OF THIS REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6666939 ↗