Advance G44145Recalls & reported problems
Cook Incorporated · ATB Advance, PTA Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002441453), received between Feb 2, 2015 and Oct 24, 2019 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture2
- Material Separation2
- Balloon rupture1
Patient problems
- Foreign Body In Patient1
- Foreign body, removal of1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 24, 2019Material Rupture; Material Separation; Material Rupture; Material Separation
AS REPORTED, DURING A FISTULAGRAM INVOLVING A (B)(6) FEMALE PATIENT, WEIGHING APPROXIMATELY (B)(6) A ATB ADVANCE BALLOON CATHETER RUPTURED AND SEPARATED INSIDE THE PATIENT. REPORTEDLY, THE BALLOON WAS DELIVERED OVER A COOK WIRE GUIDE AND WAS INFLATED ONCE IN THE LEFT INNOMINATE VEIN USING A COOK INFLATION DEVICE AND OMNIPAQUE 300 CONTRAST. THE PATIENT'S ANATOMY WAS REPORTED TO BE ANGULATED, WITH A 90-DEGREE ANGLE REPORTED. THE BALLOON WAS INFLATED TO 5 ATMOSPHERES, VERY SLOWLY, AND THE CUSTOMER NOTICED THE CENTER OF THE BALLOON WAS SMALLER THAN THE ENDS. THE CUSTOMER WAS REPORTEDLY WORKING IN THE INNOMINATE VEIN AND THE SUPERIOR VENA CAVA AT THE TIME OF INFLATION, AND THE BALLOON WAS INFLATED NEXT TO A PRE-EXISTING, UNKNOWN MANUFACTURER'S STENT. THE CUSTOMER FELT A POP, AT WHICH TIME THE DEVICE IS THOUGHT TO HAVE RUPTURED CIRCUMFERENTIALLY, AT 5 ATMOSPHERES. THE BALLOON WAS PULLED BACK, OVER THE COOK WIRE GUIDE AND WITH ANOTHER MANUFACTURER'S SHEATH, AND THE BALLOON WAS OBSERVED TO BE SHEARED OFF. USING ULTRASOUND, THE TIP OF THE DEVICE WAS CONFIRMED TO BE IN THE PATIENT'S LEFT ARM. THE USER ATTEMPTED TO REMOVE THE DEVICE WITH ANOTHER MANUFACTURER'S SNARE; HOWEVER, THE SEPARATED PO
Patient problems: Foreign Body In Patient
MAUDE report 9234719 ↗MalfunctionApr 11, 2018Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE THE BACK HUB BROKE/CRACKED OFF OF THE ATB ADVANCE BALLOON CATHETER AFTER ATTACHING THE INFLATION DEVICE. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT USING ANOTHER ATB ADVANCE BALLOON CATHETER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7420785 ↗InjuryApr 1, 2016Material Rupture; Material Rupture
DURING A BILATERAL PTA OF THE JUGULAR VEIN, THE BALLOON BURST DURING INFLATION AND HAD TO BE REMOVED FROM PATIENT. THE INCIDENT WITH THE BALLOON OCCURRED ON THE LEFT SIDE. BEFOREHAND, THEY WORKED ON THE RIGHT SIDE WITH A SECOND BALLOON WITHOUT ANY PROBLEMS (THE USER FACILITY ALWAYS TAKES A NEW BALLOON FOR EACH SIDE). NORMAL PLACEMENT OF THE ATB BALLOON OVER A WIRE GUIDE THROUGH A 8 FR INTRODUCER. THE BALLOON WAS INFLATED WITH 5 ATM. THEN THE USER FACILITY SAW THAT THEY HAD LOST A PART OF THE BALLOON. THEY WITHDREW EVERYTHING INSIDE THE SHEATH AND WITHDREW THE SHEATH AS WELL. AFTERWARDS, THEY CUT THE SHEATH TO CHECK IF THE REST OF THE BALLOON WAS INSIDE THE SHEATH, BUT IT WAS NOT. THE PATIENT WAS TRANSFERRED TO (B)(6) HOSPITAL. HERE THE REST OF THE BALLOON WAS FOUND INSIDE THE LUNG. THE PIECE WAS TAKEN OUT WITH A SNARE. THE PHYSICIANS OF THE HOSPITAL BELIEVE THAT THEY HAVE FOUND EVERYTHING. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT'S CONDITION WAS NORMAL DURING THE WHOLE PROCEDURE THERE WERE NO SIGNS IN THE EKG.
Patient problems: No Code Available
MAUDE report 5541621 ↗InjuryFeb 2, 2015Balloon rupture
DURING A PTA OF SUBCLAVIAN, THE PHYSICIAN INFLATED THE DEVICE TWICE WITH NO INCIDENT. ON THE THIRD CIRCUMFERENTIAL RUPTURE, ALL FRAGMENTS WERE ABLE TO BE REMOVED. THERE WAS NO HARM TO THE PATIENT.
Patient problems: Foreign body, removal of
MAUDE report 4491336 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page