Cook Incorporated · Black Silicone Filiform Double Pigtail Ureteral Stent Set · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002464384), received between Mar 30, 2023 and Mar 10, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Difficult to Advance2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2025Break; Break

    IT WAS REPORTED THAT THE 'BLACK SILICONE FILIFORM DOUBLE PIGTAIL URETERAL STENT' BROKE DURING REMOVAL. AS REPORTED, THREE STENTS FROM THE SAME LOT BROKE DURING REMOVAL OVER A SPAN OF TWO DAYS. REPORTEDLY, NO PATIENTS WERE HARMED. IT WAS NOTED THAT SURGEONS HAVE BEEN USING 6F STENTS, LARGER THAN 5F STENTS THEY ARE BEING USED TO. IT IS BELIEVED THAT THE ISSUE MAY BE OCCURRING WHEN ATTEMPTING TO REMOVE THE 6F STENT THROUGH A SHEATH INTENDED FOR USE WITH 5F STENTS. AS REPORTED, NO SECTION OF DEVICE REMAINS INSIDE PATIENT'S BODY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Insufficient Information

    MAUDE report 21558968 ↗
  • MalfunctionJan 20, 2025Break; Break

    ADDITIONAL INFORMATION RECEIVED 20JAN2025: THE STENT WAS INITIALLY PLACED FOR TREATMENT OF MALIGNANT COMPRESSION OF THE URETERS SECONDARY TO METASTATIC PROSTATE CANCER. THE STENT WAS BEING REMOVED FOR A ROUTINE STENT EXCHANGE. THE STENT WAS NOT BEING MONITORED BUT WAS EXCHANGED AT 8 MONTHS. THE RIGHT STENT BROKE DISTAL TO THE URETERAL ORIFICE AND WAS REMOVED WITH A CYSTOSCOPE AND STENT GRASPER DURING THE SAME PROCEDURE.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21190959 ↗
  • MalfunctionNov 13, 2024Break; Break

    IT WAS REPORTED THE BLACK SILICONE FILIFORM DOUBLE PIGTAIL URETERAL STENT BOTTOM CURL OF THE STENT BROKE OFF AS IT WAS BEING ADVANCED OVER THE WIRE BEFORE THAT PART OF THE DEVICE ENTERED THE PATIENT DURING URETEROSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER LIKE-DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20675486 ↗
  • MalfunctionMar 30, 2023Difficult to Advance; Difficult to Advance

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16648016 ↗
  • MalfunctionMar 30, 2023Difficult to Advance; Difficult to Advance

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16643477 ↗

Other Stent, ureteral devices

All Stent, ureteral devices →