Cook Spectrum G47825Recalls & reported problems

Cook Incorporated · Spectrum, Central Venous Catheter Set Minocycline/Rifampin Antibiotic Impregnated Triple Lumen Polyurethane · Class 2 · Implant

All models of Cook Spectrum (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
21
Deaths reported
0
Injuries reported
9
Recalls naming it
1

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00827002478251), received between Jun 26, 2017 and Nov 12, 2025 — classified as 9 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 5, 2026Class IIOpen, Classified

    Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

    Z-1788-2026 · Root cause: Incorrect or no expiration date · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak8
  • Crack6
  • Break2
  • Fracture2
  • Leak/Splash2
  • Obstruction of Flow2
  • Unraveled Material2
  • Complete Blockage1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information4
  • No Consequences Or Impact To Patient4
  • No Code Available3
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • Increased Level of Consciousness1
  • No Patient Involvement1

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 12, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT THE BLUE LUMEN OF A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER WAS FOUND TO BE OCCLUDED PRIOR TO AN UNKNOWN PROCEDURE FOR A MALE PATIENT. DURING TESTING PRIOR TO INSERTION, THE NURSE NOTED THAT THE BLUE LUMEN OF A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER COULD NOT BE FLUSHED, AS RESISTANCE WAS FELT WHEN ATTEMPTING TO INJECT FLUID. A NEW SET WAS OPENED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533883 ↗
  • InjuryAug 4, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE CATHETER FROM A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER SET WAS LEAKING. UPON FLUSHING THE RED LUMEN WITH SALINE, THE NURSE NOTICED A SIGNIFICANT LEAK FROM THE INSERTION SITE. TO VERIFY THE LEAK WAS FROM THE CATHETER AND NOT THE PATIENT TISSUES, A SMALL AMOUNT OF PROPOFOL WAS INJECTED INTO THE RED LUMEN. THE WHITE COLORED PROPOFOL LIQUID WAS THEN NOTED TO LEAK FROM THE INSERTION SITE. THE CATHETER WAS REMOVED AND REPLACED IMMEDIATELY. AFTER REMOVAL, THE DEVICE WAS INSPECTED BY THE CUSTOMER. UPON FLUSHING THE WHITE AND BLUE LUMENS, THE CATHETER WAS NOTED TO FUNCTION NORMALLY WITH NO LEAKS. WHEN THE RED LUMEN WAS FLUSHED, LEAKING WAS NOTED FROM AN APPROXIMATE 1 CM LONG "HAIRLINE" CRACK AT THE MOST PROXIMAL PORTION OF THE CATHETER ("WHICH WOULD HAVE BEEN RIGHT ABOUT 5 MM INTO THE PATIENT'S SKIN"). THE SITE REPORTED THAT THE RED LUMEN HAD BEEN UTILIZED FOR INFUSIONS OF ANTIBIOTICS, A MAGNESIUM INFUSION, AND BALANCED ELECTROLYTE SOLUTION. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22702482 ↗
  • MalfunctionJan 13, 2025Unraveled Material; Unraveled Material

    IT WAS REPORTED THAT THE GUIDE WIRE OF A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER SET WAS DISCOVERED TO BE KINKED AFTER REMOVAL FROM THE 78-YEAR-OLD MALE PATIENT. THE PATIENT DID NOT HAVE TORTUOUS ANATOMY. DURING THE INSERTION OF THE WIRE GUIDE THERE WAS NO DIFFICULTY ADVANCING INTO THE PATIENT. HOWEVER, IT WAS REPORTED THAT THE WIRE GUIDE "FELT BLOCKED AND STUCK." THE WIRE GUIDE WAS REMOVED FROM THE PATIENT WITHOUT DIFFICULTY. UPON REMOVAL FROM THE PATIENT, IT WAS DISCOVERED THE GUIDE WIRE WAS KINKED. THE WIRE GUIDE WAS NOT MANIPULATED OR WITHDRAWN THROUGH THE NEEDLE. A NEW DEVICE "CVP" WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE WAS RETURNED TO COOK FOR INVESTIGATION ON (B)(6)2025. DURING THE PRELIMINARY INVESTIGATION OF THE RETURNED WIRE GUIDE, IT WAS DISCOVERED THAT THE WIRE GUIDE SAFETY WIRE BROKE AWAY FROM THE WELD CONNECTION AND RESULTED IN ELONGATION. ADDITIONALLY, TWO BENDS IN THE WIRE WERE IDENTIFIED. THE BENDS IN THE WIRE WERE INITIALLY REPORTED BY THE CUSTOMER. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21139350 ↗
  • InjuryJan 9, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IN ADDITIONAL INFORMATION RECEIVED ON 13JAN2025 AND 19JAN2025, IT WAS CONFIRMED THAT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE USER REPORTED THAT NO DAMAGE WAS OBSERVED ON THE BLUE HUB.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21118151 ↗
  • InjurySep 4, 2024Crack; Leak/Splash; Crack; Leak/Splash

    IT WAS CLARIFIED BY THE CUSTOMER ON (B)(6) 2024 THAT THE CATHETER LEAKAGE OCCURRED AT THE "END HOLE" OF THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20140770 ↗
  • MalfunctionMay 6, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information

    MAUDE report 19255841 ↗
  • MalfunctionMay 6, 2024Crack; Crack

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information

    MAUDE report 19255707 ↗
  • InjurySep 7, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    ADDITIONAL INFORMATION WAS PROVIDED. AS REPORTED, THE LUMENS WERE FILLED WITH SALINE SOLUTION PRIOR TO CATHETER INTRODUCTION. NO POWER INJECTION WAS USED; IT WAS NOT A CT PORT THAT CRACKED. THE LUMEN FLOW WAS NOT IMPEDED. THERE WERE NO REPORTED ISSUES WITH PATENCY. THE LINE WAS USED FOR CONTINUOUS INFUSION PRIOR TO LEAKAGE. THE LEAK OCCURRED BETWEEN THE HUB AND THE NEEDLELESS ADAPTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17700741 ↗
  • InjuryFeb 22, 2023Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Increased Level of Consciousness

    MAUDE report 16420240 ↗
  • InjuryApr 13, 2022Fluid/Blood Leak; Fracture; Leak/Splash; Obstruction of Flow; Fluid/Blood Leak; Fracture; Leak/Splash; Obstruction of Flow

    A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER WAS INSERTED IN THE NECK OF THE PATIENT ON (B)(6)2022. THE CATHETER WAS BEING USED FOR HIGH ACUITY PATIENT TREATMENT. WHEN THE LEAK WAS DISCOVERED, REMIFENTANIL AND PROPOFOL WERE BEING ADMINISTERED THROUGH THE PROXIMAL LUMEN. THE CATHETER WAS IMMEDIATELY CLAMPED AND TAPED OVER WHEN THE LEAK WAS DISCOVERED. THE INFUSION WAS TRANSFERRED TO ANOTHER PORT. THE CATHETER WAS REMOVED AND REPLACED ON (B)(6)2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14103705 ↗
  • InjuryApr 4, 2022Material Separation; Material Separation

    IT WAS REPORTED THE WIRE GUIDE IN A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER SET SEPARATED. THE DEVICE WAS REQUIRED BY A PATIENT IN CRITICAL CONDITION FOR LARGE BORE, CENTRAL INTRAVENOUS (IV) ACCESS OF THE RIGHT FEMORAL VEIN. DURING THE PLACEMENT PROCEDURE, WHEN THE USER ADVANCED THE CATHETER OVER THE WIRE GUIDE AND INTO THE PATIENT, THE WIRE GUIDE SEPARATED. AN X-RAY WAS IMMEDIATELY TAKEN WHICH SHOWED APPROXIMATELY 22-23CM OF THE WIRE GUIDE RETAINED IN THE PELVIC VASCULATURE. THE PATIENT THEN REQUIRED AN ADDITIONAL PROCEDURE PERFORMED UNDER FLUOROSCOPY TO REMOVE THE WIRE GUIDE FRAGMENT. NO OTHER ADVERSE EVENTS WERE REPORTED DUE TO THIS OCCURRENCE. IT WAS NOTED THAT THE UNRETAINED PORTION OF THE WIRE GUIDE REMAINED INSIDE THE CATHETER. THE CATHETER WAS NOT PLACE IN THE PATIENT. IT WAS ALSO NOTED THE PROCEDURE WAS UNDERTAKEN BY EXPERIENCED STAFF AND WAS DESCRIBED AS UNCOMPLICATED WITH NO DIFFICULTY OR FORCE REQUIRED.

    Patient problems: Foreign Body In Patient

    MAUDE report 13995798 ↗
  • MalfunctionJul 13, 2020Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    DURING THE INVESTIGATION, IT WAS DETERMINED THIS EVENT WAS INADVERTENTLY REPORTED, AS IT DOES NOT MEET REPORTING CRITERIA.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10264626 ↗
  • MalfunctionJul 2, 2020Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE WIRE GUIDE OF A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER SET WAS FOUND TO BE "DEFORMED" AFTER WITHDRAWAL. THE DEVICE WAS BEING PLACED IN THE PATIENT'S LEFT VENA SUBCLAVIA. RESISTANCE WAS FELT AS THE PHYSICIAN WAS WITHDRAWING THE WIRE GUIDE THROUGH THE NEEDLE. CONSEQUENTLY, THE PHYSICIAN WITHDREW BOTH THE NEEDLE AND WIRE TOGETHER. AFTER DOING SO, THE WIRE GUIDE WAS OBSERVED TO BE "DEFORMED". THE PHYSICIAN ATTEMPTED TO USE THE OTHER SIDE OF THE WIRE GUIDE BUT ENCOUNTERED THE SAME ISSUE. AS A RESULT, A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10227336 ↗
  • MalfunctionMar 18, 2020Complete Blockage; Complete Blockage

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9849610 ↗
  • MalfunctionFeb 26, 2020Fluid/Blood Leak; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Code Available

    MAUDE report 9756278 ↗
  • MalfunctionNov 4, 2019Fluid/Blood Leak; Fluid/Blood Leak

    ADDITIONAL INFORMATION STATING THAT THE FITTING OF THE DEVICE WAS CONNECTED TO AN "AUTOMATIC CHEMICAL ADDITION AND CONTROL SYSTEM" AND THAT SYSTEM IS NOT AVAILABLE FOR RETURN. THE DEVICE WAS LEFT IN PLACE FOUR HOURS AFTER THE PROCEDURE AND WAS LOCKED WITH A NEEDLELESS CLOSED SYSTEM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9272049 ↗
  • MalfunctionJul 5, 2019Unraveled Material; Unraveled Material

    NO NEW EVENT DESCRIPTION INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8765098 ↗
  • MalfunctionJun 7, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO NEW EVENT DESCRIPTION INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Patient Involvement

    MAUDE report 8680317 ↗
  • InjurySep 21, 2018Crack; Crack

    IT WAS REPORTED A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER (CVC) HUB WAS FOUND CRACKED. THE PATIENT WILL REQUIRE A LINE REPLACEMENT PROCEDURE WHEN CONDITIONS PERMIT. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Code Available

    MAUDE report 7899062 ↗
  • MalfunctionSep 27, 2017Break; Physical Resistance

    WHEN THE DOCTOR TRIED TO REMOVE THE WIRE GUIDE FROM THE DISTAL CHANNEL, THE GUIDE PRESENTED A LOT OF RESISTANCE. WHEN THE DOCTOR FINALLY REMOVED THE WIRE GUIDE IT WAS "COMPLETELY DAMAGED". THE PATIENT DIDN'T EXPERIENCE ANY ADVERSE EFFECTS AS A RESULT OF THE PRODUCT PROBLEM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6898534 ↗
  • InjuryJun 26, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED THAT A PATIENT UNDERWENT PLACEMENT OF A COOK SPECTRUM MINOCYCLINE/RIFAMPIN IMPREGNATED TRIPLE LUMEN CENTRAL VENOUS CATHETER ON (B)(6) 2017. THE CUSTOMER STATES THAT THE CENTER LUMEN OF THE TRIPLE LUMEN DEVICE WITH THE IV TUBING CONNECTED WAS DISCOVERED LYING IN THE PATIENT'S BED. THE CIRCUMSTANCES AND FORCES APPLIED TO THE DEVICE LEADING TO THE EVENT ARE NOT KNOWN. THE LINE WAS IMMEDIATELY REMOVED AND A NEW ONE INSERTED. NO FURTHER EVENT OR PATIENT DETAILS WERE PROVIDED. THE DEVICE IS REPORTEDLY AVAILABLE FOR EVALUATION, HOWEVER IT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AS OF THE DATE OF THIS REPORT.

    Patient problems: No Code Available

    MAUDE report 6667153 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →