Skip to content

Cope G48170Recalls & reported problems

Cook Incorporated · Cope Nephroureterostomy Stent Ultrathane · Class 2

All models of Cope (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002481701), received between Feb 1, 2024 and Oct 30, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE "INNER YELLOW INSERT" HUB BROKE OFF FROM THE "BLUE COMPONENT" OF AN ULTRATHANE COPE NEPHROURETEROSTOMY SET DURING AN UNSPECIFIED PROCEDURE ON A 51-YEAR-OLD MALE PATIENT. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION ON 22OCT2024, AND UPON A REVIEW OF THE PRELIMINARY DFA FINDINGS, IT WAS FOUND THAT THE HUB OF THE FLEXIBLE STIFFENER HAD SEPARATED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20569815 ↗
  • MalfunctionMar 29, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE STIFFENER OF AN ULTRATHANE COPE NEPHROURETEROSTOMY SET WAS DIFFICULT TO REMOVE FROM THE DRAINAGE CATHETER. ADDITIONALLY, THE STRING THAT FORMS THE PROXIMAL PIGTAIL BECAME ATTACHED WHEN PULLED. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19005811 ↗
  • MalfunctionFeb 1, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE STIFFENER OF AN ULTRATHANE COPE NEPHROURETEROSTOMY SET WAS DIFFICULT TO REMOVE. THE DEVICE WAS REQUIRED FOR A STENT CHANGE PROCEDURE. AFTER THE CATHETER WAS PLACED IN THE PATIENT, THE STIFFENER COULD NOT BE REMOVED. THE CATHETER WAS THEN REMOVED AND REPLACED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18623781 ↗

Other Stent, ureteral devices

All Stent, ureteral devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page