Universa G49866Recalls & reported problems
Cook Incorporated · Universa Firm Ureteral Stent Set With Hydrophilic Coating · Class 2 · Implant
All models of Universa (82)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00827002498662), received between Apr 3, 2017 and Feb 11, 2026 — classified as 3 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Crack4
- Difficult to Remove2
- Unsealed Device Packaging2
- Adverse Event Without Identified Device or Use Problem1
- Device Damaged Prior to Use1
- Torn Material1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient4
- No Code Available2
- No Patient Involvement2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Laceration(s)1
- Renal Impairment1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2026Unsealed Device Packaging
IT WAS REPORTED THAT THE USER FOUND THE PACKAGING OF A UNIVERSA FIRM URETERAL STENT SET WAS NOT SEALED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24315593 ↗MalfunctionJan 5, 2026Break
IT WAS REPORTED, ONE OF THE PIGTAILS OF A UNIVERSA FIRM URETERAL STENT SET SEPARATED FROM THE STENT. THE ISSUE WAS DISCOVERED PRIOR TO PATIENT CONTACT AND THE DEVICE WAS NOT USED ON THE PATIENT. A NEW STENT WAS USED TO COMPLETE THE UNSPECIFIED PROCEDURE IN THE URETER. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23958196 ↗MalfunctionAug 21, 2025Break; Break
ADDITIONAL INFORMATION RECEIVED: PROCEDURE WAS COMPLETED WITH A NEW DEVICE. IT IS UNKNOWN HOW THE STENT WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22862683 ↗MalfunctionJan 7, 2022Unsealed Device Packaging; Unsealed Device Packaging
AS REPORTED, THE INNER PACKAGING OF A UNIVERSA FIRM URETERAL STENT SET WAS NOT SEALED PROPERLY. THERE WAS NO PATIENT IMPACT AS THIS EVENT OCCURRED PRIOR TO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13194978 ↗MalfunctionDec 15, 2021Break; Break
ADDITIONAL INFORMATION RECEIVED ON 10FEB2022: THE STENT COMPLETELY SEPARATED ON THE NON-TETHER SIDE OF THE DEVICE. THE TETHER WAS ATTACHED TO THE STENT WHEN THE PACKAGE WAS OPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13012081 ↗InjuryMay 11, 2021Difficult to Remove; Difficult to Remove
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11807322 ↗InjuryFeb 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, A UNIVERSA FIRM URETERAL STENT WAS IMPLANTED IN A PATIENT ON (B)(6) 2020. ON (B)(6) 2021, THE PATIENT FELT UNWELL AND RETURNED TO THE HOSPITAL FOR EXAMINATION. IT WAS DISCOVERED THAT PREVIOUSLY DIAGNOSED HYDRONEPHROSIS HAD BEEN AGGRAVATED, AND THE STONE WAS ATTACHED TO THE STENT. THE PATIENT HAD NOT RETURNED TO THE HOSPITAL FOR REVIEW DURING THIS PERIOD. THE PHYSICIAN ARRANGED FOR SURGERY, REMOVED THE STENT AFTER A LASER LITHOTRIPSY, AND EXCHANGED IT FOR A NEW DEVICE. THERE WAS NO TREATMENT REQUIRED FOR THE HYDRONEPHROSIS, WHICH SPONTANEOUSLY RESOLVED FOLLOWING THE STENT EXCHANGE. THE PATIENT WAS KNOWN TO BE IN "GOOD CONDITION."
Patient problems: No Code Available; Renal Impairment
MAUDE report 11258640 ↗MalfunctionNov 6, 2020Break; Break
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10804643 ↗MalfunctionJul 10, 2020Crack; Crack
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10257045 ↗MalfunctionJul 1, 2020Crack; Crack
IT WAS REPORTED THAT PRIOR TO AN UNSPECIFIED PROCEDURE, A CRACK IN THE PIGTAIL END OF THE UNIVERSA FIRM URETERAL STENT SET WAS FOUND WHEN THE PACKAGING WAS OPENED. THE TETHER HAD NOT BEEN REMOVED FROM THE STENT. ANOTHER UNIVERSA FIRM URETERAL STENT SET WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10218404 ↗MalfunctionJan 18, 2019Crack; Crack
IT WAS INITIALLY REPORTED DURING AN ENDOSCOPIC HOSE SURGERY, THE USER OPENED THE UNIVERSA FIRM URETERAL STENT SET PACKAGE AND FOUND OUT THE PIGTAIL END WAS BROKEN. ADDITIONAL INFORMATION RECEIVED ON 17-JAN-2019, ADVISING THE PIGTAIL WAS CRACKED AND ANOTHER SAME DEVICE TYPE WAS USED TO COMPLETE THE PROCEDURE. THIS OBSERVATION WAS MADE PRIOR TO PATIENT CONTACT AND THE STENT WAS NOT USED. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED..
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8261745 ↗InjuryJul 3, 2018Difficult to Remove; Unraveled Material; Difficult to Remove; Unraveled Material
THERE HAS BEEN NO NEW INFORMATION RECEIVED SINCE THE LAST REPORT WAS SUBMITTED.
Patient problems: Laceration(s); No Code Available
MAUDE report 7657870 ↗MalfunctionApr 12, 2018Crack; Torn Material
IT WAS REPORTED THAT THE UNIVERSA FIRM URETERAL STENT WAS ALREADY CRACKED/TORN WHEN THE PACKAGE WAS OPENED. THE DEVICE DID NOT MAKE PATIENT CONTACT, AND THERE WAS NO IMPACT TO THE PATIENT OR USERS.
Patient problems: No Patient Involvement
MAUDE report 7426615 ↗MalfunctionApr 3, 2017Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED BY THE USER FACILITY THAT UPON OPENING THE PACKAGE THE TIP OF THE STENT WAS FOUND DETACHED. THIS INCIDENT OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 6454040 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page