Cook Incorporated · Central Venous Catheter Set 5 Lumen Polyurethane · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00827002515338), received between Sep 27, 2017 and Sep 4, 2024 — classified as 4 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Material Separation4
  • Unraveled Material3
  • Out-Of-Box Failure1

Patient problems

  • Foreign Body In Patient3
  • No Known Impact Or Consequence To Patient3
  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjurySep 4, 2024Break; Break

    IT WAS REPORTED THAT THE WIRE GUIDE FROM A FIVE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET SEPARATED. DURING REMOVAL OF THE SECOND DILATOR, THE WIRE GUIDE WAS CUT AND SUBSEQUENTLY SEPARATED. TEN CENTIMETERS OF THE WIRE WAS REMOVED FROM THE PATIENT, AND THE REMAINDER WAS RETAINED IN THE INTRAMUSCULAR. AS A RESULT, THE PATIENT REQUIRED IMAGING AND DEVICE REMOVAL IN THE PAVILION. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20141078 ↗
  • MalfunctionMar 11, 2024Break; Material Separation; Break; Material Separation

    ADDITIONAL INFORMATION WAS PROVIDED STATING THE WIRE WAS MANIPULATED THROUGH THE NEEDLE. THE PHYSICIAN INDICATED THAT THE WIRE GUIDE MANDRIL BROKE DURING THE DILATION PROCESS WHEN THE DILATOR PASSED OVER THE GUIDE. ALL FRAGMENTS WERE REMOVED FROM THE PATIENT DURING THE INITIAL PLACEMENT PROCEDURE. IT IS UNKNOWN IF THE PATIENT HAD TORTUOUS ANATOMY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18878183 ↗
  • InjuryMar 11, 2024Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 18878062 ↗
  • MalfunctionNov 6, 2020Unraveled Material; Unraveled Material

    IN ADDITIONAL INFORMATION RECEIVED ON 24NOV2020, IT WAS REPORTED THAT THE NEEDLE WAS NOT WITHDRAWN THROUGH THE NEEDLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10802178 ↗
  • MalfunctionNov 6, 2020Unraveled Material; Unraveled Material

    IT WAS REPORTED THAT THERE WAS DIFFICULTY REMOVING THE WIRE GUIDE DURING INSERTION OF A FIVE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER ON (B)(6) 2020. UPON REMOVAL OF THE WIRE GUIDE, IT WAS FOUND TO BE BROKEN. A MALE PATIENT WAS ADMITTED TO THE ADULT CRITICAL PATIENT UNIT FOR ACUTE RESPIRATORY FAILURE ASSOCIATED WITH COVID-19, AND WAS REPORTED TO BE IN RESPIRATORY ISOLATION. THE INDICATION FOR PLACEMENT WAS ADMINISTRATION OF MEDICATION, HEMODERIVATIVES, AND OTHER TREATMENT. A PUNCTURE WAS MADE IN THE SUBCLAVIAN VEIN UNDER "STERILE TECHNIQUE". THE GUIDE WIRE WAS INSERTED AND AFTER DILATION, AND ATTEMPT WAS MADE TO INSERT THE CATHETER. THERE WAS DIFFICULTY EXPERIENCED INSERTING THE CATHETER, AND AT THIS POINT, THE WIRE GUIDE WAS REMOVED AND FOUND TO BE UNRAVELED. DUE TO THE POSSIBLE RISK OF INJURING THE PATIENT WITH THE BROKEN WIRE GUIDE, THE ENTIRE DEVICE WAS REMOVED FROM THE PATIENT. A DIFFERENT MANUFACTURER'S CATHETER WAS USED AT A DIFFERENT PUNCTURE SITE TO COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10802109 ↗
  • MalfunctionJul 20, 2020Unraveled Material; Unraveled Material

    ADDITIONAL INFORMATION WAS PROVIDED ON 31JUL2020. THE GUIDE WIRE WAS FOUND TO BE DAMAGED BEFORE THE DEVICE WAS TO BE PLACED IN THE LEFT SUBCLAVIAN VEIN. THE DEVICE DID NOT MAKE PATIENT CONTACT.

    Patient problems: No Patient Involvement

    MAUDE report 10297863 ↗
  • InjuryJul 20, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT THE WIRE GUIDE OF A FIVE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET BROKE. NO ADVERSE EFFECTS HAVE BEEN REPORTED. ADDITIONAL INFORMATION REGARDING PATIENT AND EVENT INFORMATION HAS BEEN REQUESTED BUT IS NOT CURRENTLY AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10297673 ↗
  • InjuryJul 20, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT THE WIRE GUIDE OF A FIVE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET SEPARATED AND THE DISTAL PORTION OF THE WIRE MIGRATED TO THE PATIENT'S HEART. THE PATIENT HAD TO HAVE AN ADDITIONAL PROCEDURE TO REMOVE THE PORTION OF THE WIRE THAT MIGRATED TO THEIR HEART. ADDITIONAL INFORMATION REGARDING PATIENT AND EVENT INFORMATION HAS BEEN REQUESTED BUT IS NOT CURRENTLY AVAILABLE.

    Patient problems: Foreign Body In Patient; No Code Available

    MAUDE report 10296253 ↗
  • MalfunctionSep 27, 2017Break; Out-Of-Box Failure

    WHEN THE PHYSICIAN WAS OPENING THE PACKAGE, HE OBSERVED THAT THE CATHETER WAS BROKEN, AND THIS OCCURRED WITH FOUR DIFFERENT LOT NUMBERS. THE PHYSICIAN MANAGED TO USE A FIFITH PRODUCT TO FINISH THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT OR ADDITIONAL PROCEDURES HAVE BEEN REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6896941 ↗
  • MalfunctionSep 27, 2017Break

    WHEN THE PHYSICIAN WAS OPENING THE PACKAGE, HE OBSERVED THAT THE CATHETER WAS BROKEN, AND THIS OCCURRED WITH FOUR DIFFERENT LOT NUMBERS. THE PHYSICIAN MANAGED TO USE A FIFTH PRODUCT TO FINISH THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT OR ADDITIONAL PROCEDURES HAVE BEEN REPORTED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6896740 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →