CXI G52543Recalls & reported problems
Cook Incorporated · CXI Support Catheter · Class 2
All models of CXI (34)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00827002525436), received between Dec 1, 2016 and Apr 18, 2025 — classified as 2 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Material Separation” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation5
- Detachment of Device or Device Component3
- Failure to Advance2
- Material Twisted/Bent2
- Appropriate Device Problem Term/Code Not Available1
- Detachment Of Device Component1
- Difficult to Remove1
- Kinked1
Patient problems
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient4
- Device Embedded In Tissue or Plaque2
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Code Available1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 18, 2025Difficult to Remove; Difficult to Remove
AS REPORTED, DURING A PROCEDURE INVOLVING ANGIOPLASTY OF THE SUPERFICIAL FEMORAL ARTERY (SFA), ANOTHER MANUFACTURER¿S 0.035-INCH WIRE BECAME STUCK IN A CXI SUPPORT CATHETER. IPSILATERAL ACCESS WAS OBTAINED IN THE COMMON FEMORAL ARTERY TO TARGET THE POPLITEAL ARTERY. THE CXI WAS ADVANCED THROUGH ANOTHER MANUFACTURER¿S 5-FRENCH SHEATH, AND THE CATHETER REPORTEDLY PASSED SOME HEAVY CALCIFICATION AS IT WAS ADVANCED INTO THE SFA. RESISTANCE WAS NOT ENCOUNTERED DURING INSERTION OR ADVANCEMENT OF THE CATHETER. WITHIN THE FIRST THIRTY MINUTES OF THE PROCEDURE, THE DECISION WAS MADE TO EXCHANGE THE WIRE FOR A HYDROPHILIC WIRE, TO CONTINUE FURTHER DOWN THE LIMB AND INTO THE POPLITEAL VESSEL. AS THE USER ATTEMPTED TO REMOVE THE WIRE FROM THE CATHETER, IT WOULD NOT COME BACK THROUGH THE CATHETER AND APPEARED TO BE STUCK; THEREFORE, THE USER CUT THE HUB OF THE CATHETER, AND THE WIRE WAS REMOVED. THE CATHETER DID NOT SEPARATE. PER THE USER, THE WIRE WAS INTACT UPON REMOVAL. THE WIRE WAS NOT ALTERED PRIOR TO USE. ANOTHER CXI CATHETER AND 0.018-INCH WIRE WERE USED TO REACH THE TARGET LESION AND SUCCESSFULLY COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21864480 ↗MalfunctionFeb 10, 2022Material Separation; Material Separation
AS REPORTED, DURING A VASCULAR ANGIOGRAM AND STENT PLACEMENT, A CXI SUPPORT CATHETER WAS PLACED VIA FEMORAL ACCESS OVER ANOTHER MANUFACTURERS GUIDE WIRE BUT WAS UNABLE TO BE ADVANCED PAST "3 TO 5 CENTIMETERS." THE WIRE AND CXI WERE REMOVED AND A PIECE WAS SHEARED OFF. IT IS UNKNOWN IF THE SHEARED PIECE IS FROM THE COOK CATHETER OR THE OTHER MANUFACTURER'S GUIDE WIRE. NO UNINTENDED PART OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. NO ADDITIONAL PROCEDURE WAS REQUIRED DUE TO THIS OCCURRENCE. NO ADVERSE EFFECT ON THE PATIENT WAS REPORTED DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13505380 ↗InjuryMar 1, 2021Material Separation; Material Separation
NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: Insufficient Information
MAUDE report 11391922 ↗MalfunctionMar 2, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING AN UNKNOWN PROCEDURE, A CXI SUPPORT CATHETER HUB SEPARATED FROM THE DEVICE. ADDITIONAL INFORMATION REGARDING EVENT DETAILS, PATIENT ANATOMY AND OUTCOME HAS BEEN REQUESTED, BUT IS NOT AVAILABLE AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9774607 ↗MalfunctionDec 17, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8170124 ↗MalfunctionSep 11, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED, THE HUB BECAME DETACHED FROM THE SHAFT OF THE CXI SUPPORT CATHETER DURING A TIBIAL PERONEAL TRUNK INTERVENTION PROCEDURE. THIS OCCURRED WHILE TRYING TO CROSS A TIGHT, "VERY" CALCIFIC LESION WHILE TORQUING THE CATHETER. A SECOND CXI SUPPORT CATHETER WAS USED WHICH ALSO COULD NOT CROSS THE LESION BUT NO HUB DETACHMENT WAS REPORTED. THE PROCEDURE WAS ABLE TO BE COMPLETED USING ANOTHER MANUFACTURER'S CATHETER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7865328 ↗MalfunctionJan 18, 2018Material Separation; Material Separation
IT WAS REPORTED THAT DURING A SUPERFICIAL FEMORAL ARTERY (SFA) INTERVENTION FOR RECANALIZATION, THE CXI SUPPORT CATHETER STRAIN RELIEF HUB SNAPPED OFF AS THE CATHETER WAS BEING REMOVED. IT WAS NOTED THAT THE PHYSICIAN PULLED THE CATHETER BY THE HUB DURING REMOVAL. CONTRALATERAL ACCESS TO THE PATIENT'S COMMON FEMORAL ARTERY WAS GAINED BY CROSSING OVER THE AORTOILIAC BIFURCATION UTILIZING A SHEATH. THE PATIENT'S VESSEL WAS NOTED TO BE CALCIFIED. ANOTHER CATHETER WAS UTILIZED TO SUCCESSFULLY COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION REGARDING EVENT DETAILS, PATIENT ANATOMY AND OUTCOME HAS BEEN REQUESTED, BUT IS NOT AVAILABLE AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7203157 ↗MalfunctionNov 2, 2017Stretched; Material Twisted/Bent
THE PHYSICIAN USED A CXI CATHETER AND THE BODY OF THE CATHETER BECAME TWISTED AND STRETCHED. THE CATHETER WAS REMOVED AND ANOTHER DEVICE USED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6995751 ↗MalfunctionNov 1, 2017Kinked; Leak/Splash; Failure to Advance; Positioning Problem
AS REPORTED TO CUSTOMER RELATIONS, INITIALLY, THE DOCTOR WAS ABLE TO PASS A WIRE ALL THE WAY THROUGH THE CATHETER. AT THIS POINT THE WIRE WAS REMOVED AND A SHOT OF CONTRAST WAS INJECTED THRU THE CATHETER. THE PHYSICIAN WENT TO PUT THE WIRE (G52543) BACK THRU THE CATHETER IT WOULD NOT PASS THRU. THERE WAS NO KINK VISIBLE AND THE ACT WAS ABOVE 300. THEY USED A PILOT 150 WIRE AND THE CONTRAST USED WAS OMNIPAQUE 300. EVERYTHING WAS REMOVED AND THEY WERE ABLE TO PASS THE STENOSIS WITH THE G52543; HOWEVER ONCE THEY WERE PASSED THE STENOSIS THEY WERE NOT ABLE TO ADVANCE THE WIRE ANY FURTHER BECAUSE THERE WAS A KINK IN THE CATHETER. THIS KINK IS RIGHT WHERE ONE OF THE BLACK BANDS IS ON THE CATHETER. THIS DEVICE WAS PULLED OUT OF THE BODY AND OBSERVED BY THE PHYSICIAN. IT WAS KINKED SO BAD THAT THE BRAIDING WAS EXPOSED. THE TWO ABOVE DEVICES WERE UTILIZED THRU THE ACCESS POINT OF THE LEFT RADIAL ARTERY. THE PHYSICIAN WAS USING A G56232 IN THE RIGHT COMMON FEMORAL AND THE VALVE LEAKED THRU THE ENTIRE CASE. THIS SHEATH WAS USED UNTIL THE PROCEDURE WAS COMPLETED AS THIS WAS A SLOW LEAK. THERE WERE NO ADDITIONAL PROCEDURES; NO ADVERSE EFFECTS REPORTED AND NO DEVICE BREAKAGE OR SEPARATION REPORT
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6992699 ↗MalfunctionOct 31, 2017Detachment Of Device Component
CXI CATHETER WAS USED TO CROSS A DISTAL POPLITEAL LESION. AS THE PHYSICIAN WAS TORQUING THE HUB OF THE CATHETER, THE HUB DETACHED. THE CATHETER WAS REMOVED AND ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE AND ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6991738 ↗MalfunctionOct 30, 2017Failure to Advance; Material Twisted/Bent
THE CXI SUPPORT CATHETER WAS USED FOR A LEFT LEG ARTERIOGRAM INTERVENTION. THE CATHETER WAS OVER A WIRE GUIDE, SUBINTIMAL, AND USED TO CROSS A VERY TIGHT LONG LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE DOCTOR TRIED TO ROTATE THE CATHETER MULTIPLE TIMES TO REDIRECT THE TIP, CAUSING THE MIDDLE OF THE CATHETER TO TWIST TO THE POINT WHERE THE CATHETER WOULD NO LONGER ADVANCE OVER THE WIRE GUIDE. THE CATHETER WAS REMOVED AND ANOTHER CATHETER WAS USED TO COMPLETE THE PROCEDURE WITH NO HARM TO THE PATIENT. DEVICE IMAGING IDENTIFIED TWISTING IN THE CATHETER TO THE POINT OF COMPROMISE IN MATERIAL INTEGRITY.A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6987201 ↗MalfunctionAug 14, 2017Material Separation; Material Separation
A CXI SUPPORT CATHETER WAS USED IN AN 'UP & OVER RIGHT LEG ANGIOPLASTY & WOULD DEBRIDEMENT" PROCEDURE. IT WAS REPORTED THAT THE DEVICE SNAPPED OFF WHILE TRAVELING THROUGH STENOSIS. 15CM OF CXI WAS LEFT INSIDE THE PATIENT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE AND, ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6791815 ↗InjuryDec 1, 2016Material Separation; Appropriate Device Problem Term/Code Not Available; Material Separation; Appropriate Device Problem Term/Code Not Available
AS REPORTED TO CUSTOMER RELATIONS: "WHILE TRYING TO CANULATE THE RIGHT RENAL AND GAIN ACCESS, THE TIP OF THE CATHETER BROKE AND REMAINED IN THE RIGHT RENAL ARTERY. IT NEVER GOT RETRIEVED. THE PHYSICIAN PUT A PLUG IN THE RIGHT FENESTRATION TO BLOCK THE RIGHT RENAL BECAUSE OF ENDOLEAK CREATED WITH THE FACT THAT THE RIGHT RENAL WAS NOT LEFT PATENT. PATIENT WILL LOOSE HER RIGHT." ADDITIONAL INFORMATION WAS REQUESTED FOR THE CUSTOMER.
Patient problems: Device Embedded In Tissue or Plaque; No Code Available
MAUDE report 6137617 ↗
FDA clearance and classification
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CXI G52543
Has the CXI G52543 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002525436), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CXI G52543?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Dec 1, 2016 and Apr 18, 2025. In those reports, the most common specific device problems are “Material Separation” (5), “Detachment of Device or Device Component” (3) and “Failure to Advance” (2); the most common patient problem is “Device Embedded In Tissue or Plaque” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CXI G52543?
None of the reports are classified as deaths, 2 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the CXI G52543, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand CXI, model G52543). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page