Skip to content

CXI G52544Recalls & reported problems

Cook Incorporated · CXI Support Catheter · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002525443), received between Sep 18, 2017 and May 22, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Contamination /Decontamination Problem1
  • Delamination1
  • Device Contamination with Chemical or Other Material1
  • Difficult to Advance1
  • Kinked1
  • Material Frayed1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement2
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2025Break; Break

    AS REPORTED, UPON OPENING THE DEVICE PACKAGE PRIOR TO USE FOR AN UNKNOWN PROCEDURE, THE USER NOTED THAT THE TIP OF A CXI SUPPORT CATHETER WAS BENT AND ¿IT LOOKED LIKE THE COILED HAD COME LOOSE.¿ THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069756 ↗
  • MalfunctionJun 14, 2024Break; Break

    AS REPORTED, DURING AN UNSPECIFIED VASCULAR PROCEDURE, A CXI SUPPORT CATHETER SEPARATED UPON REMOVAL OF THE DEVICE. PER THE REPORTER, THE CATHETER BROKE IN MULTIPLE LOCATIONS ALONG THE SHAFT AND AT THE HUB. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE COMPLAINT DEVICE. ALL CATHETER FRAGMENTS WERE REMOVED MANUALLY OVER THE UNKNOWN WIRE GUIDE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19537492 ↗
  • MalfunctionOct 26, 2021Material Frayed; Material Frayed

    NO ADDITIONAL INFORMATION HAS BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12697556 ↗
  • MalfunctionJan 16, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Patient Involvement

    MAUDE report 9595751 ↗
  • InjuryJul 11, 2018Difficult to Advance; Difficult to Advance

    IT WAS REPORTED, DURING A PROCEDURE TO TREAT THE CHRONIC TOTAL OCCLUSION (CTO) OF THE PATIENT'S SUPERFICIAL FEMORAL ARTERY (SFA), THE PHYSICIAN COULD NOT ADVANCE THE CXI SUPPORT CATHETER THROUGH THE TARGET LESION. THE COMPLAINT DEVICE WAS REMOVED FROM THE PATIENT'S ANATOMY AT THE DISCRETION OF THE PHYSICIAN BECAUSE IT WAS "FELT SOMETHING WAS WRONG". SUBSEQUENTLY, ANOTHER MANUFACTURER'S MICRO CATHETER WAS SELECTED TO CONTINUE THE PROCEDURE; HOWEVER, BECAUSE A WIRE GUIDE COULD NOT ADVANCE THROUGH THE TRUE LUMEN, THE PROCEDURE WAS STOPPED WITH NO FURTHER TREATMENT. THE PROCEDURE WAS RESCHEDULED FOR THE END OF (B)(6) 2018 TO BE COMPLETED AT ANOTHER FACILITY BY A DIFFERENT PHYSICIAN. ACCORDING TO THE INITIAL REPORTER, THERE HAVE BEEN NO PATIENT ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Code Available

    MAUDE report 7677213 ↗
  • MalfunctionNov 22, 2017Delamination; Delamination

    IT WAS REPORTED THAT DURING THE PERIPHERAL VASCULAR PROCEDURE, WHITE PIECES OF THE CXI SUPPORT CATHETER COATING WERE COMING OFF. THE DELAMINATED MATERIAL DID NOT ENTER THE PATIENT'S ANATOMY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7053988 ↗
  • MalfunctionNov 17, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    DURING INCOMING INSPECTION OF THE DEVICE AT THE DISTRIBUTER, IT WAS OBSERVED THAT A PIECE OF AN UNIDENTIFIED PARTICLE WAS INSIDE THE PRIMARY PACKAGE. THIS DEVICE WAS NOT USED AT A FACILITY. NO PATIENT CONTACT WAS MADE. THERE WERE NO ADVERSE EVENTS REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 7042068 ↗
  • MalfunctionOct 31, 2017Kinked; Stretched

    DURING ATTEMPT TO EXCHANGE A WIRE, THE WIRE WOULD KINK WHILE TRYING TO PASS THROUGH THE CATHETER. THE DEVICE WAS ALSO NOTED TO HAVE STRETCHED. A SECOND CATHETER WAS USED AND THE SAME DIFFICULTY OCCURRED. THE PROCEDURE WAS ENDED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE AND ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6990649 ↗
  • MalfunctionSep 18, 2017Break; Material Separation; Break; Material Separation

    A CXI SUPPORT CATHETER WAS USED DURING MULTIPLE PROCEDURES. PER A VOLUNTARY MEDWATCH (# (B)(4)), IT WAS REPORTED THAT UPON REMOVAL OF THE CATHETER OVER AN AMPLATZ WIRE, THE LUER-LOCK HUB BROKE OFF OF THE DEVICE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6872010 ↗

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.