CXI G52544Recalls & reported problems
Cook Incorporated · CXI Support Catheter · Class 2
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002525443), received between Sep 18, 2017 and May 22, 2025 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Contamination /Decontamination Problem1
- Delamination1
- Device Contamination with Chemical or Other Material1
- Difficult to Advance1
- Kinked1
- Material Frayed1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
- No Patient Involvement2
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2025Break; Break
AS REPORTED, UPON OPENING THE DEVICE PACKAGE PRIOR TO USE FOR AN UNKNOWN PROCEDURE, THE USER NOTED THAT THE TIP OF A CXI SUPPORT CATHETER WAS BENT AND ¿IT LOOKED LIKE THE COILED HAD COME LOOSE.¿ THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22069756 ↗MalfunctionJun 14, 2024Break; Break
AS REPORTED, DURING AN UNSPECIFIED VASCULAR PROCEDURE, A CXI SUPPORT CATHETER SEPARATED UPON REMOVAL OF THE DEVICE. PER THE REPORTER, THE CATHETER BROKE IN MULTIPLE LOCATIONS ALONG THE SHAFT AND AT THE HUB. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE COMPLAINT DEVICE. ALL CATHETER FRAGMENTS WERE REMOVED MANUALLY OVER THE UNKNOWN WIRE GUIDE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19537492 ↗MalfunctionOct 26, 2021Material Frayed; Material Frayed
NO ADDITIONAL INFORMATION HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12697556 ↗MalfunctionJan 16, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Patient Involvement
MAUDE report 9595751 ↗InjuryJul 11, 2018Difficult to Advance; Difficult to Advance
IT WAS REPORTED, DURING A PROCEDURE TO TREAT THE CHRONIC TOTAL OCCLUSION (CTO) OF THE PATIENT'S SUPERFICIAL FEMORAL ARTERY (SFA), THE PHYSICIAN COULD NOT ADVANCE THE CXI SUPPORT CATHETER THROUGH THE TARGET LESION. THE COMPLAINT DEVICE WAS REMOVED FROM THE PATIENT'S ANATOMY AT THE DISCRETION OF THE PHYSICIAN BECAUSE IT WAS "FELT SOMETHING WAS WRONG". SUBSEQUENTLY, ANOTHER MANUFACTURER'S MICRO CATHETER WAS SELECTED TO CONTINUE THE PROCEDURE; HOWEVER, BECAUSE A WIRE GUIDE COULD NOT ADVANCE THROUGH THE TRUE LUMEN, THE PROCEDURE WAS STOPPED WITH NO FURTHER TREATMENT. THE PROCEDURE WAS RESCHEDULED FOR THE END OF (B)(6) 2018 TO BE COMPLETED AT ANOTHER FACILITY BY A DIFFERENT PHYSICIAN. ACCORDING TO THE INITIAL REPORTER, THERE HAVE BEEN NO PATIENT ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 7677213 ↗MalfunctionNov 22, 2017Delamination; Delamination
IT WAS REPORTED THAT DURING THE PERIPHERAL VASCULAR PROCEDURE, WHITE PIECES OF THE CXI SUPPORT CATHETER COATING WERE COMING OFF. THE DELAMINATED MATERIAL DID NOT ENTER THE PATIENT'S ANATOMY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7053988 ↗MalfunctionNov 17, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
DURING INCOMING INSPECTION OF THE DEVICE AT THE DISTRIBUTER, IT WAS OBSERVED THAT A PIECE OF AN UNIDENTIFIED PARTICLE WAS INSIDE THE PRIMARY PACKAGE. THIS DEVICE WAS NOT USED AT A FACILITY. NO PATIENT CONTACT WAS MADE. THERE WERE NO ADVERSE EVENTS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 7042068 ↗MalfunctionOct 31, 2017Kinked; Stretched
DURING ATTEMPT TO EXCHANGE A WIRE, THE WIRE WOULD KINK WHILE TRYING TO PASS THROUGH THE CATHETER. THE DEVICE WAS ALSO NOTED TO HAVE STRETCHED. A SECOND CATHETER WAS USED AND THE SAME DIFFICULTY OCCURRED. THE PROCEDURE WAS ENDED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE AND ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6990649 ↗MalfunctionSep 18, 2017Break; Material Separation; Break; Material Separation
A CXI SUPPORT CATHETER WAS USED DURING MULTIPLE PROCEDURES. PER A VOLUNTARY MEDWATCH (# (B)(4)), IT WAS REPORTED THAT UPON REMOVAL OF THE CATHETER OVER AN AMPLATZ WIRE, THE LUER-LOCK HUB BROKE OFF OF THE DEVICE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6872010 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.