Aprima G52963Recalls & reported problems
Cook Incorporated · Aprima Access Set Nonvascular Introducer Set Featuring Transitionless Technology · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002529632), received between Dec 5, 2014 and Dec 5, 2014 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation1
Patient problems
- Foreign body, removal of1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 5, 2014Material Separation
DURING A BILIARY DRAINAGE INSERTION ON A FEMALE PATIENT, THE PHYSICIAN HAD INSERTED THE 22 G CHIBA NEEDLE PERCUTANEOUSLY AND HE PUT THE GUIDE WIRE INTO IT IN ORDER TO PERFORM THE SELDINGER TECHNIQUE. HE TWISTED THE GUIDE WIRE IN ORDER TO ENTER IN A BRANCH OF THE BILIARY TREE, BUT AFTER SEVERAL ATTEMPTS, THE TIP OF THE WIRE SNAPPED, REMAINING WITHIN THE BODY. THE DOCTOR LEFT THE TIP INSIDE IN ORDER NOT TO TAKE FURTHER RISKS. THE DOCTOR IS UNSURE ABOUT LOT HE USED BUT HE ALSO USED ANOTHER LOT (1820334-2014-00607) DURING THE PROCEDURE. A SECTION OF THE DEVICE DID REMAIN INSIDE THE PATIENT'S BODY. AFTER 4 DAYS, THE PHYSICIAN PERFORMED ANOTHER PROCEDURE AS THE PATIENT NEEDED TO RECEIVE A BILIARY DRAINAGE INSERTION DUE TO JAUNDICE. DURING THIS PROCEDURE AS THE PHYSICIAN GAINED ACCESS TO BILIARY TREE HE RETRIEVED THE GUIDE WIRE PIECE WITH A LOOP RETRIEVAL CATHETER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Foreign body, removal of
MAUDE report 4335999 ↗
Other ACCESSORIES, CATHETER devices
- Cook G00219Cook Incorporated · ACCESSORIES, CATHETER103 reports
- Discofix® 456020B. Braun Medical Inc. · ACCESSORIES, CATHETER60 reports
- Extension Set 456025B. Braun Medical Inc. · ACCESSORIES, CATHETER1 recall0 reports
- Centeze Centesis Catheter DRC-002-06Galt Medical Corp. · Accessories, Catheter1 recall0 reports
- Centeze Centesis Catheter DRC-002-05Galt Medical Corp. · Accessories, Catheter1 recall0 reports
- IV Administration Set 490062B. Braun Medical Inc. · ACCESSORIES, CATHETER1 recall0 reports
- Neff G10544Cook Incorporated · ACCESSORIES, CATHETER45 reports
- Cook G06299Cook Incorporated · ACCESSORIES, CATHETER18 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.