Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002529649), received between Mar 19, 2015 and Mar 19, 2015 — classified as 1 injury.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2015Material Separation

    WHEN ATTEMPTING TO PLACE A KIDNEY DRAINAGE CATHETER ON (B)(6) 2015, THE PHYSICIAN INSERTED THE ACCESS NEEDLE, THEN INSERTED THE COPE MANDRIL WIRE FROM AP RIMA SET. AFTER REMOVING THE NEEDLE AND MANIPULATING THE WIRE TO TRY TO GET PAST THE STRICTURE, THE PHYSICIAN REMOVED THE WIRE AND THE TECH NOTICED THAT THE FLOPPY TIP WAS NO LONGER PRESENT AFTER TAKING FURTHER IMAGES, THE TIP OF THE WIRE CAN BE SEEN INSIDE THE PATIENT'S KIDNEY THE PHYSICIAN STILL TRIED TO GAIN ACCESS TO PLACE THE DRAINAGE BY OPENING UP 3 SUBSEQUENT A PRIMA SETS, BUT WAS UNSUCCESSFUL IN GETTING PAST THE STRICTURE. IT WAS THEN DECIDED TO WAIT AND LET PATIENT HEAL UP FROM TODAY'S PROCEDURE BEFORE DECIDING THE NEXT COURSE TRY AGAIN TO PLACE DRAIN PERCUTANEOUSLY OR GO TO PARTIAL NEPHRECTOMY AS WAS PREVIOUSLY DISCUSSED AS A POTENTIAL COURSE OF TREATMENT FOR THE PATIENT. THERE WAS NOT AN ATTEMPT TO REMOVE THE TIP OF THE WIRE ON THIS DAY, BUT IF THEY DECIDE TO ATTEMPT PERCUTANEOUS ACCESS- THEY WILL TRY TO USE FORCEPS/SNARE TO REMOVE WIRE TIP. TECH STATED THAT THE PHYSICIAN DID NOT MOVE THE WIRE IN AND OUT OF THE NEEDLE DURING THE MANIPULATION. FOLLOW UP PROCEDURE: AFTER A SECOND ATTEMPT ON (B)(6) 2015 TO ACCESS THE KIDNE

    Patient problems: Foreign Body In Patient

    MAUDE report 4626996 ↗