Coda G53025Recalls & reported problems
Cook Incorporated · Coda, Balloon Catheter · Class 2
All models of Coda (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002530256), received between Feb 20, 2017 and Nov 12, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Deflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 12, 2025Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23532355 ↗MalfunctionJun 18, 2025Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22288097 ↗MalfunctionJan 30, 2019Material Rupture; Material Rupture
IT WAS REPORTED THE PHYSICIAN PLACED A STENT GRAFT FROM ZONE 3. THE APPROACH WAS OBTAINED FROM THE RIGHT CFA. THE PHYSICIAN USED THE COMPLAINT (B)(4) FOR BALLOON TOUCH-UPS. HE DID BALLOON TOUCH-UPS AT BOTH OF THE PROXIMAL SIDE AND THE DISTAL SIDE OF THE STENT GRAFT. AFTER THAT, A TYPE IA ENDOLEAK WAS CONFIRMED BY ANGIOGRAPHY ON THE COMPETITOR'S GRAFT. THEN, THE PHYSICIAN ATTEMPTED BALLOON TOUCH-UPS AT THE PROXIMAL SIDE OF THE STENT GRAFT AGAIN. HOWEVER, THE BALLOON RUPTURED. ANOTHER (B)(4) WAS USED INSTEAD TO COMPLETE THE PROCEDURE. THE PROCEDURE ENDED SUCCESSFULLY. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8288840 ↗MalfunctionJun 11, 2018Deflation Problem
IT WAS REPORTED THAT THE BALLOON OF THE CODA BALLOON CATHETER WOULD NOT DEFLATE AFTER BEING INFLATED INSIDE A STENT GRAFT FOR AN AORTIC ANEURYSM REPAIR PROCEDURE. THERE WERE NO INJURIES OR ADDITIONAL PROCEDURES REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7589104 ↗MalfunctionFeb 20, 2017Material Rupture; Material Rupture
DURING A THORACIC ENDOVASCULAR AORTIC REPAIR PROCEDURE, A CODA BALLOON CATHETER WAS INFLATED WITH TRIPLY DILUTED CONTRAST MEDIA WITH APPROPRIATE PRESSURE AMOUNT BUT RUPTURED INSIDE THE STENT GRAFT AT THIRD INFLATION. NO ENDOLEAK WAS FOUND, SO THE PROCEDURE WAS COMPLETED WITHOUT USING ANOTHER BALLOON CATHETER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6342965 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗