Skip to content

TriForce G56412Recalls & reported problems

Cook Incorporated · TriForce Peripheral Crossing Set · Class 2

All models of TriForce (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002564121), received between Apr 9, 2024 and Apr 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    AS REPORTED, DURING AN UNSPECIFIED PROCEDURE INVOLVING A PATIENT WITH CENTRAL VENOUS STENOSIS, THE TIP OF A TRIFORCE PERIPHERAL CROSSING SET'S SHEATH SPLIT. ACCESS WAS OBTAINED IN THE RIGHT UPPER ARM TO TARGET THE BRACHIAL VEIN, AND THE ANATOMY WAS VERY CALCIFIED. THE HYDROPHILIC COATING WAS ACTIVATED, AND THE SHEATH WAS ADVANCED WITH THE DILATOR, PRIOR TO CATHETER ADVANCEMENT. THE DEVICE WAS USED TO CROSS A CHRONIC OCCLUSION. RESISTANCE WAS NOT FELT DURING INSERTION OR REMOVAL OF THE DEVICE; HOWEVER, WHEN THE DEVICE WAS REMOVED, THE USER NOTED THAT THE SHEATH TIP HAD SPLIT. A POWER INJECTOR WAS NOT USED. THE VESSELS WERE NOT DAMAGED. ANOTHER MANUFACTURER'S GRAFT WAS USED TO COMPLETE THE PROCEDURE. PER THE REPORTER, THE GRAFT WAS NOT PLACED DUE TO THE SPLIT IN THE SHEATH TIP, BUT WAS ALREADY PLANNED IF THE OCCLUSION WAS UNABLE TO BE CROSSED, WHICH IT WASN'T. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19072482 ↗

FDA clearance and classification

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TriForce G56412

Has the TriForce G56412 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002564121), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TriForce G56412?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 9, 2024. That report lists device problem “Material Split, Cut or Torn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TriForce G56412?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the TriForce G56412, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand TriForce, model G56412). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates