TriForce G56415Recalls & reported problems
Cook Incorporated · TriForce Peripheral Crossing Set · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002564152), received between May 16, 2017 and Sep 3, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Separation1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2025Break; Break
AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE TIP OF A TRIFORCE PERIPHERAL CROSSING SET BROKE. PER THE REPORTER, THE TIP APPEARS TO BE PARTIALLY ATTACHED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED. UPON RETURN AND INITIAL EVALUATION OF THE COMPLAINT DEVICE, THE TIP OF THE SHEATH WAS SEPARATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22955897 ↗MalfunctionFeb 4, 2020Material Separation; Material Separation
AS REPORTED, DURING A RECANALIZATION PROCEDURE OF THE FEMORAL VEIN INVOLVING A PATIENT WITH DEEP VEIN THROMBOSIS, A TRIFORCE PERIPHERAL CROSSING SET SHEATH KINKED UPON ADVANCEMENT INTO A CHRONIC CLOT. ACCESS WAS OBTAINED IN THE BACK OF THE LEG. THE ANATOMY WAS REPORTED TO BE CALCIFIED. THE PROCEDURE WAS COMPLETED WITH ANOTHER MANUFACTURER'S SHEATH, USING THE CATHETER INCLUDED WITH THE TRIFORCE PERIPHERAL CROSSING SET. UPON RETURN AND INITIAL EVALUATION OF THE DEVICE ON 03FEB2020, THE SHEATH WAS NOTED TO BE PARTIALLY SEPARATED IN TWO PLACES. THE TIP WAS ALSO NOTED TO BE DAMAGED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9667157 ↗MalfunctionMay 16, 2017Peeled/Delaminated
AS THE FACILITY WAS PREPPING THE DEVICE BEFORE THE PROCEDURE, IT WAS NOTED THAT THE SHEATH'S MATERIAL WAS PEELING OFF TOWARDS THE END OF THE DEVICE. THE FACILITY OPENED A NEW PRODUCT (UNKNOWN MAKE AND MODEL) TO COMPLETE THE INTENDED PROCEDURE. THE COMPLAINT DEVICE DID NOT MAKE PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 6568649 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.