TriForce G56416Recalls & reported problems
Cook Incorporated · TriForce Peripheral Crossing Set · Class 2
All models of TriForce (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002564169), received between Oct 23, 2019 and Nov 14, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Split, Cut or Torn1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2024Unraveled Material; Unraveled Material
ADDITIONAL INFORMATION WAS RECEIVED 13NOV2024 AND INADVERTENTLY OMITTED FROM THE INITIAL REPORT. THE BREAK/UNRAVELING WAS NOTED AFTER SUCCESSFUL USE AND REMOVAL OF THE DEVICE, PER THE INSTRUCTIONS FOR USE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT INCIDENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT. NO ADDITIONAL INFORMATION WILL BE PROVIDED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20682832 ↗MalfunctionJan 15, 2024Break; Break
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18512665 ↗InjuryMay 24, 2023Break; Break
AS REPORTED, DURING THE MIDDLE OF A PROCEDURE INVOLVING RECONSTRUCTION OF THE INFERIOR VENA CAVA, VIA ACCESS IN THE FEMORAL ARTERY, THE TIP BROKE OFF OF A TRIFORCE PERIPHERAL CROSSING SET'S SHEATH. THE SEPARATED TIP WAS BALLOONED INTO THE SHEATH AND REMOVED FROM THE PATIENT SUCCESSFULLY. SCISSORS WERE THEN USED TO REMOVE THE ENCAPSULATED BROKEN TIP. THERE HAS BEEN NO REPORT OF ANY ADVERSE EFFECTS TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16992546 ↗MalfunctionOct 23, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
ADDITIONAL INFORMATION WAS RECEIVED ON 05NOV2019 NOTING THAT THE PROCEDURE BEING PERFORMED WAS A CHRONIC ILIOFEMORAL OUTFLOW OBSTRUCTION. THE FLEXOR SHEATH SPLIT WHILE TRYING TO CROSS THE LESION. REPORTEDLY, THE PHYSICIAN NOTED THAT THERE WAS ONE LOCATION WHERE HE WAS TRYING TO CROSS AND "YOU COULD SEE THE FLEXOR TIP GET 'HUNG UP' ON THE LESION". ADDITIONALLY, RESISTANCE DURING INSERTION OR RESISTANCE WAS NOT NOTED. PRIOR TO USE, THE SHEATH WAS ADEQUATELY PREPPED AND FLUSHED, AND THE HYDROPHILIC COATING WAS ACTIVATED. ACCORDING TO THE INITIAL REPORTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY OTHER COMPLICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9225286 ↗
FDA clearance and classification
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TriForce G56416
Has the TriForce G56416 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002564169), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TriForce G56416?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 23, 2019 and Nov 14, 2024. In those reports, the most common specific device problems are “Break” (2), “Material Split, Cut or Torn” (1) and “Unraveled Material” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TriForce G56416?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TriForce G56416, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand TriForce, model G56416). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page