Skip to content

CXI G56552Recalls & reported problems

Cook Incorporated · CXI Support Catheter · Class 2

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00827002565524), received between Nov 5, 2020 and Jan 16, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Device Contaminated During Manufacture or Shipping2
  • Material Separation2
  • Contamination /Decontamination Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient2
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Break

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE TIP OF A CXI SUPPORT CATHETER SEPARATED PRIOR TO PATIENT CONTACT AS IT WAS INSERTED INTO A SHEATH AND ADVANCED OVER A WIRE GUIDE. A NEW CXI CATHETER WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093397 ↗
  • InjuryDec 30, 2025Break; Break

    AS REPORTED, DURING A LOWER EXTREMITY ANGIOGRAM, THE TIP OF A CXI SUPPORT CATHETER SEPARATED INSIDE OF THE PATIENT. ACCESS WAS OBTAINED IN THE RIGHT FEMORAL ARTERY TO TARGET THE SUPERFICIAL FEMORAL ARTERY AND PROFUNDA. ANOTHER MANUFACTURER¿S SHEATH AND ANOTHER MANUFACTURER¿S WIRE WERE USED DURING THE PROCEDURE. THE HYDROPHILIC COATING WAS REPORTEDLY ACTIVATED. RESISTANCE WAS ENCOUNTERED DURING INSERTION/ADVANCEMENT OF THE CATHETER, AS THE LESION WAS CALCIFIED. DURING REMOVAL OF THE OTHER MANUFACTURER¿S 0.014-INCH WIRE GUIDE, THE TIP OF THE CATHETER SEPARATED ¿FROM THE DISTAL MARKER FORWARD¿. PER THE REPORTER, IT IS UNCLEAR IF THE WIRE WAS PULLED FIRST OR IF THE CATHETER WAS PULLED WITH THE WIRE INSIDE. A POWER INJECTOR WAS NOT USED WITH THE CATHETER. THE USER WAS UNABLE TO CROSS THE LESION; THEREFORE, THE PROCEDURE WAS NOT COMPLETED. THE SEPARATED FRAGMENT REMAINS IN THE PATIENT, WITHIN A LATERAL BRANCH OF THE PROFUNDA. PER THE REPORTER, THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23926754 ↗
  • InjuryNov 4, 2025Break; Break

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23468789 ↗
  • MalfunctionDec 13, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED BY A DISTRIBUTOR, FOREIGN MATTER WAS NOTED IN A CXI SUPPORT CATHETER'S STERILE PACKAGE. THERE WERE NO ADVERSE EFFECTS TO ANY PATIENT. THE DEVICE WAS NOT USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20929649 ↗
  • MalfunctionJul 14, 2022Material Separation; Material Separation

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, A WIRE CAME OUT OF A HOLE IN THE SIDE OF A CXI SUPPORT CATHETER. UPON RETURN AND INITIAL EVALUATION OF THE DEVICE, THE DISTAL TIP WAS NOTED TO BE PARTIALLY SEPARATED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15012885 ↗
  • MalfunctionDec 16, 2021Material Separation; Material Separation

    NO NEW PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13017440 ↗
  • MalfunctionAug 20, 2021Contamination /Decontamination Problem; Device Contaminated During Manufacture or Shipping; Contamination /Decontamination Problem; Device Contaminated During Manufacture or Shipping

    AS REPORTED, AN UNKNOWN SUBSTANCE WAS ATTACHED TO A CXI SUPPORT CATHETER PRIOR TO USE. A PHOTO PROVIDED BY THE CUSTOMER SHOWS A DARK, STRING-LIKE SUBSTANCE, WHICH APPEARS TO BE ATTACHED TO THE HUB OF THE DEVICE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12349029 ↗
  • MalfunctionNov 5, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10795613 ↗