Echelon 190-5091-150Recalls & reported problems
Micro Therapeutics, Inc. · 10 Micro Catheter;90 Degree Tip Shape · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00836462002357), received between May 15, 2026 and May 15, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K093750); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT A COIL ENCOUNTERED RESISTANCE AT THE HUB OF A ECHELON CATHETER. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A COIL EMBOLIZATION WITH AN ANTERIOR CEREBRAL ARTERY (ACA) ACCESS VESSEL OF 2.5MM. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE COIL GOT CAUGHT AT THE HUB DURING INSERTION WHICH OCCURRED TWICE. THE HYDRO SOFT, AND TARGET TETRA SIZE WAS UNKNOWN, SO IT WAS DETERMINED TO BE A PROBLEM WITH THE CATHETER. THERE WAS RESISTANCE AT THE HUB OF THE CATHETER. THE SHEATH TIP WAS FIRMLY POSITIONED AT THE HUB. THE CATHETER WAS COLLECTED, CHANGED TO SL10 THEN THE PROCEDURE CONTINUED. DEVICES WERE PREPARED AS INDICATED IN THE PACKAGE INSERT. NO PATIENT COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED. THE CONCOMITANT COIL WAS NOT A MEDTRONIC PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25186471 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K093750. One submission often covers many models, so check the product and lot details in each record.
- Z-1556-2022 · Jun 30, 2022 · Open, ClassifiedEchelon -14, 45 Tip Shape, REF 145-5092150, sterile EO, Rx Only
- Z-1557-2022 · Jun 30, 2022 · Open, ClassifiedEchelon -14 Micro Catheter, REF: 105-5092-150, sterile EO, Rx Only
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page