Skip to content

HyperForm 104-4470 (UDI 00836462004184)Recalls & reported problems

Micro Therapeutics, Inc. · Occlusion Balloon System · Class 2

All models of HyperForm (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00836462004184), received between May 27, 2020 and Jan 20, 2026 — classified as 1 death, 10 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel6
  • Inflation Problem5
  • Leak/Splash3
  • Deflation Problem1
  • Device Damaged by Another Device1
  • Packaging Problem1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Death1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Burst Container or Vessel; Inflation Problem

    ADDITIONAL INFORMATION RECEIVED. THE PATIENT WAS UNDERGOING TREATMENT FOR A SACCULAR RIGHT CA ANEURYSM, APPROXIMATELY 5 MM, WITH NORMAL VESSEL TORTUOSITY. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU, INCLUDING INSPECTION AND HYDRATION. THE CAUSE OF THE BALLOON LEAK WAS NOT DETERMINED. DURING THE TEST SET UP, THE GUIDEWIRE TIP WAS ADVANCED APPROXIMATELY 3 CM OUT OF THE CATHETER TIP, AND THE CONTRAST RATIO USED WAS 50%. THE BALLOON WAS INFLATED ONCE OUTSIDE THE BODY AND TWICE INSIDE, USING A MEDALLION 1ML SYRINGE AT A STEADY INJECTION RATE.THE LOCATION OF THE BALLOON LEAK WAS UNKNOWN, AND THE PHYSICIAN DID NOT SHAPE THE GUIDEWIRE TIP. DURING THE PROCEDURE, GUIDEWIRE MANIPULATION WAS NORMAL, AND AN XPEDION GUIDEWIRE WAS USED, WITH THE TIP ADVANCED ABOUT 3 MM OUT OF THE CATHETER TIP DURING INFLATION. THE CONTRAST RATIO WAS 50% AND INJECTION WAS STEADY. BOTH THE BALLOON AND XPEDION GUIDEWIRE WERE HYDRATED AS INDICATED IN THE IFU. BALLOON DEFLATION WAS PERFORMED USING A 1ML MEDALLION TO APPLY NEGATIVE PRESSURE, WITH THE GUIDEWIRE ADVANCED APPROXIMATELY 3 CM OUT OF THE CATHETER TIP. IT IS UNKNOWN WHETHER THE CATHETER AND GUIDEWIRE WERE INSPECTED FOR BALLOON PERFORMANCE OR CLOT PRESENC

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116038 ↗
  • MalfunctionNov 20, 2025Burst Container or Vessel; Inflation Problem; Packaging Problem; Burst Container or Vessel; Inflation Problem; Packaging Problem

    MEDTRONIC RECEIVED A REPORT THAT A HOLE WAS FOUND IN THE HYPERFORM BALLOON DURING PREPARATION. IMMEDIATELY AFTER AIR WAS RELEASED ACCORDING TO THE INSTRUCTIONS FOR USE AT THE TIME OF OPENING, THE BALLOON DID NOT EXPAND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23610403 ↗
  • MalfunctionJun 24, 2025Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem

    MEDTRONIC RECEIVED A REPORT THAT BECAUSE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS PERFORMED AFTER PIPELINE (PED2-450-20) PLACEMENT, THE BALLOON WAS NOT CONTRAST-ENHANCED WHEN THE HYPERFORM WAS INSERTED AND EXPANDED IN ACCORDANCE WITH THE PIPELINE'S STENT DISTAL. THERE WAS NO RESISTANCE FELT DURING INFLATION, AND UPON REMOVAL, AN ATTEMPT WAS MADE TO EXPAND THE BALLOON OUTSIDE THE BODY, BUT CONTRAST MEDIUM WAS LEAKING FROM THE BALLOON. PREPARATION WAS PERFORMED AS PER THE PACKAGE INSERT PRIOR TO INSERTION, BUT THERE WERE NO PROBLEMS WITH THE EXPANSION AND CONTRACTION. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22320105 ↗
  • MalfunctionOct 11, 2024Inflation Problem; Leak/Splash; Physical Resistance/Sticking; Inflation Problem; Leak/Splash; Physical Resistance/Sticking

    ADDITIONAL INFORMATION WAS RECEIVED. GUIDEWIRE MANIPULATION WAS USED WITHIN THE RANGE OF ABOUT 1-8 CM. THE GUIDEWIRE THAT WAS USED WITH THE BALLOON WAS NON-MEDTRONIC. DURING THE INFLATION, THE GUIDEWIRE TIP WAS ADVANCED OUT OF THE CATHETER TIP WITHIN ABOUT 8 CM. THE PHYSICIAN SHAPED THE GUIDEWIRE TIP INTO A J-SHAPE. CONTRAST RATIO THAT WAS USED WAS 50/50. INJECTION RATE WAS STEADY. THE BALLOON WAS FLUSHED AS INDICATED IN THE IFU. IT IS UNKNOWN IF THE GUIDEWIRE HYDRATED FOR MORE THAN 30 SECONDS. A NORMAL IN-DEF WITH A SYRINGE WAS DONE TO DEFLATE THE BALLOON. IT WAS CONFIRMED THAT THE CATHETER AND GUIDEWIRE WAS REMOVED OUT OF THE BODY, BUT IT SEEMS THAT NO BLOOD CLOT ADHESION HAS BEEN CONFIRMED. IT IS UNKNOWN IF BLOOD ENTERED THE CATHETER LUMEN. IT SEEMED THAT IT WAS A LITTLE DIFFICULT TO GUIDE DURING THE CATHETER OPERATION, BUT A KINK WAS NOT CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20428917 ↗
  • MalfunctionSep 23, 2024Burst Container or Vessel; Burst Container or Vessel

    ADDITIONAL INFORMATION RECEIVED THAT WHEN ASKED IF THE PHYSICIAN TEST THE BALLOON PRIOR TO USE, THE REP STATED THE PHYSICIAN PERFORMED AS PROCEDURE. THE BALLOON PERFORMED AS INTENDED DURING TESTING. THE GUIDEWIRE TIP WAS ADVANCED OUT OF THE CATHETER TIP 3MM TO 5 MM DURING THE INFLATION. THE CONTRAST RATIO WAS A DILUTION RATE, IT'S HALF. THE PHYSICIAN SLIGHTLY SHAPED THE GUIDEWIRE TIP. THE BALLOON GOT INFLATED AND DEFLATED ZERO TIMES. THE INJECTION RATE WAS STEADY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20288382 ↗
  • MalfunctionMar 12, 2024Device Damaged by Another Device; Device Damaged by Another Device

    MEDTRONIC RECEIVED A REPORT THAT AN EVENT IN WHICH 10 GUIDEWIRE WAS PROTRUDING FROM THE HYPERFORM BALLOON DURING PREPARATION AT 10 GUIDEWIRE DURING TREATMENT. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18888355 ↗
  • MalfunctionAug 28, 2023Burst Container or Vessel; Burst Container or Vessel

    MEDTRONIC RECEIVED A REPORT THAT DURING PREPARATION, IT WAS FOUND THAT THE CONTRAST AGENT WAS LEAKING FROM THE HYPERFORM BALLOON SO IT WAS REPLACED WITH A NEW ONE FOR PRECAUTION. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17638638 ↗
  • MalfunctionNov 29, 2022Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    MEDTRONIC RECEIVED A REPORT THAT THE BALLOON WAS OCCLUDED. THE HYPERFORM47 WAS SELECTED FOR IC-4MM BAC CASE. IT WAS CONFIRMED THAT T HERE WAS NO DAMAGE TO THE OUTER BOX, SO IT WAS OPENED. THE PREPARATION AND AIR REMOVAL WERE PERFORMED AS PER THE PACKAGE INSERT, BUT DURING THE TEST INFLATION, THE BALLOON DID NOT EXPAND NORMALLY. THE SUPPLIED X-PEDION WAS USED AS PER THE PACKAGE INSERT. AS PER THE PACKAGE INSERT, IT WAS SHIELDED WITHIN 10 CM OF X-PEDION TIP PART. HOWEVER, WHEN THE BALLOON WAS CLOSELY LOOKED, IT WAS CONFIRMED THAT THE CONTRAST AGENT LEAKED FROM THE BALLOON. AFTER THAT, A NEW NON-MEDTRONIC BALLOON/TRANSFORM PRODUCT WAS OPENED AND THE PROCEDURE WAS PERFORMED SUCCESSFULLY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15877120 ↗
  • MalfunctionOct 31, 2022Leak/Splash; Leak/Splash

    ADDITIONAL INFORMATION WAS RECEIVED. THE DEVICE WAS PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU) (INSPECTION ,HYDRATION, ETC.). THE CAUSE OF THE CONTRAST LEAK WAS NOT DETERMINED. THE DETACHABLE INJECTION PORT WAS NOT USED. THE PHYSICIAN TESTED THE BALLOON PRIOR TO USE. IT IS UNKNOWN IF THE BALLOON PERFORMED AS INTENDED DURING TESTING. IT IS UNKNOWN HOW FAR THE GUIDEWIRE TIP WAS ADVANCED OUT OF THE CATHETER TIP DURING THE INFLATION. THE BALLOON LEAK WAS LOCATED IN THE MIDDLE SEGMENT. THE CONTRAST RATIO WAS 50:50. THE PHYSICIAN DID NOT SHAPE THE GUIDEWIRE. IT IS UNKNOWN HOW MANY TIMES THE BALLOON INFLATED AND DEFLATED. THE INJECTION RATE WAS STEADY. IT IS UNKNOWN WHAT TYPE OF SYRINGE WAS USED DURING INFLATION/DEFLATION OF THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15705480 ↗
  • MalfunctionMar 8, 2022Burst Container or Vessel; Burst Container or Vessel

    MEDTRONIC RECEIVED A REPORT THAT WHEN THE HYPERFORM BALLOON WAS INFLATED FOR PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA), RUPTURE WAS NOTED AND THE PRODUCT WAS REMOVED OUT OF PATIENT'S BODY. THE PRODUCT WAS DISPOSED BY THE FACILITY BECAUSE THE PATIENT WAS SUSPECTED TO HAVE AN INFECTION. THE BALLOON WAS REPLACED, AND THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13689524 ↗
  • DeathMay 27, 2020Deflation Problem; Deflation Problem

    ADDITIONAL INFORMATION: THE DEVICES WILL BE RETURNED IN THE FUTURE. THERE ARE NO IMAGES TO PROVIDE. A 1ML SYRINGE WAS USED. THE BALLOON WAS FLUSHED AS PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO FRICTION OR DIFFICULTY DURING THE INFLATION. THE ISSUE THE COIL DETACHED WITHIN THE CATHETER AND THEN CAME OUT OF THE CATHETER. AS THE BALLON FAILED TO DEFLATE DURING PREPARATION, IT WAS NOT USED DURING THE PROCEDURE AND DID NOT CONTRIBUTE TO PATIENT DEATH.

    Patient problems: Death; No Known Impact Or Consequence To Patient

    MAUDE report 10092530 ↗

FDA clearance and classification

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HyperForm 104-4470

Has the HyperForm 104-4470 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00836462004184), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HyperForm 104-4470?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between May 27, 2020 and Jan 20, 2026. In those reports, the most common specific device problems are “Burst Container or Vessel” (6), “Inflation Problem” (5) and “Leak/Splash” (3); the most common patient problem is “Death” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HyperForm 104-4470?

1 report is classified as a death, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the HyperForm 104-4470, and is it still sold?

The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperForm, model 104-4470). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, intravascular occluding, temporary devices

All Catheter, intravascular occluding, temporary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page