Skip to content

Pipeline PED-450-18 (UDI 00836462015531)Recalls & reported problems

Micro Therapeutics, Inc. · Flex Embolization Device · Class 3 · Implant

All models of PIPELINE (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00836462015531), received between Nov 11, 2019 and Nov 11, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Activation Failure1
  • Physical Resistance/Sticking1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2019Activation Failure; Physical Resistance/Sticking; Activation Failure; Physical Resistance/Sticking

    MEDTRONIC RECEIVED INFORMATION THAT THE PIPELINE FLEX DISTAL BRAID GOT STRETCHED AS THE PHYSICIAN WAS ATTEMPTING TO DEPLOY IT AND IT COULD NOT OPEN. WHILE PHYSICIAN TRIED TO OPEN THE DISTAL BRAID OF THE DEVICE, IT WAS GOT STRETCHED. IT WAS NOTED THAT THE PIPELINE WAS LOCKED UP IN THE DISTAL PART OF THE MICROCATHETER. THE PHYSICIAN RELEASED THE SLACK IN AN ATTEMPT TO RESOLVE THE PROBLEM, BUT IT DID NOT RESOLVE THE ISSUE. THERE WAS NO DAMAGE OBSERVED ON THE MICROCATHETER OR THE PUSHWIRE. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THE EVENT. THE CATHETER FLUSHED CONTINUOUSLY WITH HEPARINIZED SALINE. THE PATIENT WAS UNDERGOING EMBOLIZATION TREATMENT OF AN UNRUPTURED BLISTER ANEURYSM MEASURING 1.5MM X 1MM LOCATED IN THE OPHTHALMIC SEGMENT BLISTER OF THE INTERNAL CAROTID ARTERY (ICA). THE DISTAL AND PROXIMAL LANDING ZONE WAS 3.4MM X 4.5MM. THE VASCULATURE WAS MODERATE IN TORTUOSITY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9303110 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.

  • Z-1245-2025 · Jan 30, 2025 · Open, Classified
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
  • Z-1246-2025 · Jan 30, 2025 · Open, Classified
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
  • Z-2402-2021 · Jul 13, 2021 · Open, Classified
    Pipeline Flex Embolization Device
  • Z-2403-2021 · Jul 13, 2021 · Open, Classified
    EV3 Pipeline Flex Embolization Device with Flex Shield Technology
  • Z-1527-2020 · Feb 14, 2020 · Terminated
    ev3 Pipeline Flex Embolization Device. For neurological endovascular use.
  • Z-0308-2017 · Oct 5, 2016 · Terminated
    Alligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
  • Z-0309-2017 · Oct 5, 2016 · Terminated
    Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
  • Z-0310-2017 · Oct 5, 2016 · Terminated
    X-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
  • Z-0307-2017 · Oct 5, 2016 · Terminated
    Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
  • Z-0120-2016 · Sep 4, 2015 · Terminated
    -Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
  • Z-1453-2014 · Apr 1, 2014 · Terminated
    Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
  • Z-2045-2013 · Jul 19, 2013 · Terminated
    Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck

Other Intracranial aneurysm flow diverter devices

All Intracranial aneurysm flow diverter devices →