Pipeline PED-500-25 (UDI 00836462015913)Recalls & reported problems
Micro Therapeutics, Inc. · Flex Embolization Device · Class 3 · Implant
All models of PIPELINE (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00836462015913), received between May 13, 2019 and May 11, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Activation Failure2
- Break2
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2021Activation Failure; Activation Failure
ADDITIONAL INFORMATION RECEIVED INDICATED THERE WERE NO IMAGES AVAILABLE FOR REVIEW. THERE WAS NO DAMAGE OBSERVED TO THE PUSHWIRE, AND WHEN THE DEVICE WAS OUTSIDE OF THE BODY, THE MID PORTION WAS STILL NOT OPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11805336 ↗MalfunctionDec 21, 2020Activation Failure; Physical Resistance/Sticking; Activation Failure; Physical Resistance/Sticking
MEDTRONIC RECEIVED INFORMATION REGARDING FOUR PIPELINE DEVICES THAT FAILED TO OPEN. THE PATIENT WAS UNDERGOING A FLOW DIVERSION PROCEDURE TO TREAT AN UNRUPTURED SACCULAR ANEURYSM OF THE CAVERNOUS SEGMENT OF AN INTERNAL CAROTID ARTERY (ICU). THE ANEURYSM MAX DIAMETER WAS 14MM AND THE NECK DIAMETER WAS 8MM. THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. DUAL ANTIPLATELET TREATMENT (DAPT) WAS ADMINISTERED WITH NOTED PRU LEVEL OF 160. IT WAS REPORTED THAT AN ATTEMPT WAS FIRST MADE WITH A PIPELINE (MODEL: PED-500-25, LOT: A718991). THE PIPELINE WAS ADVANCED THROUGH THE PHENOM-27 150CM MICROCATHETER WITH SOME RESISTANCE. DURING THE INITIAL DEPLOYMENT WITH THE MIDDLE SECTION OF THE PIPELINE POSITIONED IN A VESSEL BEND, THE DISTAL END OF THE PIPELINE WOULD NOT OPEN. MORE THAN 50% OF THE PIPELINE HAD BEEN DEPLOYED. NO ADDITIONAL STEPS WERE TAKEN TO TRY AND OPEN THIS PIPELINE. THE PIPELINE WAS RESHEATHED AND REMOVED WITH THROUGH THE PHENOM CATHETER. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED RELATED TO THE EVENT. POST-PROCEDURE ANGIOGRAPHY REVEALED NO ISSUE. THE ANEURYSM REMAINED UNRUPTURED AND THERE WAS NO EVIDENCE OF DAMAGE FROM THE DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11047419 ↗MalfunctionJan 13, 2020Break; Break
MEDTRONIC RECEIVED REPORT THAT DEPLOYMENT ISSUES WERE ENCOUNTERED WITH THE PIPELINE FLEX AND THAT THE PIPELINE WAS BROKEN. THE PATIENT WAS REPORTED TO HAVE BEEN TREATED FOR A POST TRAUMATIC AUTOMOBILE ACCIDENT THAT RESULTED IN A LEFT, PROXIMAL, INTERNAL CAROTID ARTERY (ICA) PSEUDOANEURYSM /FISTULA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9580320 ↗MalfunctionMay 13, 2019Break; Break
MEDTRONIC RECEIVED REPORT THAT THE PIPELINE EXPERIENCED RESISTANCE WHEN ADVANCED THROUGH THE MICROCATHETER. THE PIPELINE WAS DIFFICULT TO PLACES. THE PIPELINE WAS DEPLOYED FROM C2 TO CAVERNOUS SINUS SITE AND IT DID NOT DEPLOY WELL. THE DEVICE WAS ATTEMPTED TO BE RESHEATHED, HOWEVER, THE DEVICE ENCOUNTERED RESISTANCE AND THE DEVICE WAS UNABLE TO BE RESHEATHED. THE PIPELINE FELL (MIGRATED) FROM THE DISTAL DEPLOYMENT POSITION, WHICH MADE IT DIFFICULT TO PLACE. THE PIPELINE WAS REMOVED FROM THE PATIENT. UPON ASSESSMENT OF THE REMOVED DEVICE, THE DISTAL PART OF THE STENT WAS BROKEN, AND THE MICROCATHETER CORRUGATED (ACCORDIONED). THE ANEURYSM WAS IN THE LEFT INTERNAL CAROTID ARTERY - FROM C3 TO C4. THE ANEURYSM WAS UNRUPTURED, AMORPHOUS AND THE NECK WIDTH WAS 10MM OR MORE. THE LANDING ZONE WAS ABOUT 3.87 MM DISTALLY AND ABOUT 5.22 MM PROXIMAL. THE ANATOMY WAS SEVERELY TORTUOUS AND ARTERIOSCLEROSIS, AND IT WAS VERY DIFFICULT TO OPERATE PIPELINE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE SURGICAL TIME WAS EXTENDED BY LESS THAN 30 MIN. A CONTINUOUS FLUSH WAS USED DURING THE PROCEDURE. THE DEVICES WERE EACH PREPARED PER THE INSTRUCTIONS FOR USE (IFU). THE PATIENT HARM: NO. VESSEL
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8603911 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-2402-2021 · Jul 13, 2021 · Open, ClassifiedPipeline Flex Embolization Device
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page