Reform Pedicle Screw System 39-PA-8570Recalls & reported problems

Precision Spine, Inc. · 8.5 x 70mm Polyaxial Screw · Class 2 · Implant

All models of Reform Pedicle Screw System (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019900515), received between May 19, 2025 and Dec 31, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fracture1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 31, 2025Fracture; Fracture

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2025, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. SURGEON NOTED THE ONE OF THE SCREWS WAS INCORRECTLY POSITION AND DURING ATTEMPTS TO REMOVE IT, THE TIP OF THE RETENTION BONE-SCREWDRIVER (39-SP-0601) BROKE. A SMALL ROD WAS INSERTED INTO THE SCREW AND THE ENTIRE SCREW WAS SUCCESSFULLY REMOVED USING A STRONG ROD HOLDER. THE SURGEON THEN ATTEMPTED TO REPOSITION AND REINSERT THE SAME SCREW. HOWEVER, IT REMAINED INCORRECTLY PLACED AND REQUIRED REMOVAL. DURING THE REMOVAL ATTEMPT THE HEAD OF THE REFORM 8.5 X 70MM POLYAXIAL PEDICLE SCREW (39-PA-8570) FRACTURED. THE SCREW WAS ULTIMATELY EXTRACTED USING A ROD. THE SCREW HEAD BROKE COMPLETELY OFF DURING THE FINAL STAGE OF REMOVAL. WHILE INSERTING A NEW SCREW, THE TIP OF THE POLYAXIAL DRIVER ASSEMBLY REFORM PEDICLE SCREW SYSTEM (39-SP-0730) BROKE. A ROD WAS USED TO REMOVE THE SCREW AS THE DRIVER TIP WAS STUCK IN THE SCREW HEAD, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT FURTHER ISSUE. THERE WAS A FIVE (5) MINUTE DELAY TO THE PROCEDURE AS A RESULT OF THE ISSUES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23936080 ↗
  • InjuryMay 19, 2025Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE WAS PERFORMED ON (B)(6) 2025, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. WHILE ATTEMPTING TO INSERT AN UNKNOWN SCREW, THE TIP OF THET20 POLYAXIAL DRIVER REFORM PEDICLE SCREW SYSTEM (39-SP-0730) BROKE AND SUBSEQUENTLY WHEN TRYING TO REMOVE THE SCREW THE DRIVER BROKE AGAIN. SUBSEQUENTLY, WHEN ATTEMPTING TO INSERT THE SECOND SCREW USING THE RETENTION BONE-SCREWDRIVER (39-SP-0601) THE SCREW BROKE. ADDITIONAL DETAILS HAVE BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 22045601 ↗

Other Orthosis, Spinal Pedicle Fixation devices

All Orthosis, Spinal Pedicle Fixation devices →