Reform Pedicle Screw System 39-PA-8570Recalls & reported problems
Precision Spine, Inc. · 8.5 x 70mm Polyaxial Screw · Class 2 · Implant
All models of Reform Pedicle Screw System (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019900515), received between May 19, 2025 and Dec 31, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Fracture1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2025Fracture; Fracture
IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2025, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. SURGEON NOTED THE ONE OF THE SCREWS WAS INCORRECTLY POSITION AND DURING ATTEMPTS TO REMOVE IT, THE TIP OF THE RETENTION BONE-SCREWDRIVER (39-SP-0601) BROKE. A SMALL ROD WAS INSERTED INTO THE SCREW AND THE ENTIRE SCREW WAS SUCCESSFULLY REMOVED USING A STRONG ROD HOLDER. THE SURGEON THEN ATTEMPTED TO REPOSITION AND REINSERT THE SAME SCREW. HOWEVER, IT REMAINED INCORRECTLY PLACED AND REQUIRED REMOVAL. DURING THE REMOVAL ATTEMPT THE HEAD OF THE REFORM 8.5 X 70MM POLYAXIAL PEDICLE SCREW (39-PA-8570) FRACTURED. THE SCREW WAS ULTIMATELY EXTRACTED USING A ROD. THE SCREW HEAD BROKE COMPLETELY OFF DURING THE FINAL STAGE OF REMOVAL. WHILE INSERTING A NEW SCREW, THE TIP OF THE POLYAXIAL DRIVER ASSEMBLY REFORM PEDICLE SCREW SYSTEM (39-SP-0730) BROKE. A ROD WAS USED TO REMOVE THE SCREW AS THE DRIVER TIP WAS STUCK IN THE SCREW HEAD, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT FURTHER ISSUE. THERE WAS A FIVE (5) MINUTE DELAY TO THE PROCEDURE AS A RESULT OF THE ISSUES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936080 ↗InjuryMay 19, 2025Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE WAS PERFORMED ON (B)(6) 2025, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. WHILE ATTEMPTING TO INSERT AN UNKNOWN SCREW, THE TIP OF THET20 POLYAXIAL DRIVER REFORM PEDICLE SCREW SYSTEM (39-SP-0730) BROKE AND SUBSEQUENTLY WHEN TRYING TO REMOVE THE SCREW THE DRIVER BROKE AGAIN. SUBSEQUENTLY, WHEN ATTEMPTING TO INSERT THE SECOND SCREW USING THE RETENTION BONE-SCREWDRIVER (39-SP-0601) THE SCREW BROKE. ADDITIONAL DETAILS HAVE BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 22045601 ↗
Other Orthosis, Spinal Pedicle Fixation devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page