Skip to content

Slimplicity Anterior Cervical Plate System SDV4014Recalls & reported problems

Precision Spine, Inc. · 4.0mm x 14mm SD Variable Screw · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019905206), received between Sep 10, 2019 and Sep 10, 2019 — classified as 2 injuries.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Migration or Expulsion of Device2

Patient problems

  • No Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2019Migration or Expulsion of Device; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT UNDERWENT INITIAL IMPLANTATION ON (B)(6) 2019 UTILIZING THE SIMPLICITY ANTERIOR CERVICAL PLATE SYSTEM. REVISION WAS PERFORMED ON (B)(6) 2019 TO ADDRESS BACK OUT OF TWO (2) 4.0MM X 14MM SD VARIABLE SCREWS (SDV4014) FROM THE C7 SEGMENT. ALL HARDWARE WAS REMOVED AND REPLACED. RADIOGRAPH PROVIDED SHOWS TWO SCREWS BACKED OUT OF THE BONE BUT NOT BACKED OUT OF THE PLATE. FOLLOWING RECEIPT OF EXPLANTED PRODUCT IT WAS IDENTIFIED THAT NONE OF THE LOCKING TABS ON THE PLATE ARE BROKEN.

    Patient problems: No Information

    MAUDE report 8983991 ↗
  • InjurySep 10, 2019Migration or Expulsion of Device; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT UNDERWENT INITIAL IMPLANTATION ON (B)(6) 2019 UTILIZING THE SIMPLICITY ANTERIOR CERVICAL PLATE SYSTEM. REVISION WAS PERFORMED ON (B)(6) 2019 TO ADDRESS BACK OUT OF TWO (2) 4.0MM X 14MM SD VARIABLE SCREWS (SDV4014) FROM THE C7 SEGMENT. ALL HARDWARE WAS REMOVED AND REPLACED. RADIOGRAPH PROVIDED SHOWS TWO SCREWS BACKED OUT OF THE BONE BUT NOT BACKED OUT OF THE PLATE. FOLLOWING RECEIPT OF EXPLANTED PRODUCT IT WAS IDENTIFIED THAT NONE OF THE LOCKING TABS ON THE PLATE ARE BROKEN.

    Patient problems: No Information

    MAUDE report 8983978 ↗

Other Appliance, Fixation, Spinal Intervertebral Body devices

All Appliance, Fixation, Spinal Intervertebral Body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page