Reform MC Screw System 59-SP-0601Recalls & reported problems

Precision Spine, Inc. · Adjustment Screw Driver · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019927987), received between Feb 9, 2023 and Jul 23, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Entrapment of Device1
  • Fracture1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2024 UTILIZING THE REFORM MC SCREW SYSTEM. DURING THE PROCEDURE THE T25, ADJUSTABLE DRIVER; REFORM MC SCREW SYSTEM (59-SP-0601) TIP BROKE. THE BROKEN TIP WAS REMOVED FROM THE SCREW AND THE PROCEDURE WAS COMPLETED UTILIZING OTHER INTRUMENTATION READILY AVAILABLE. THERE WAS NO PATIENT INJURY, BUT A FIFTEEN (15) MINUTE DELAY TO THE PROCEDURE RESULTED FROM THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19810827 ↗
  • InjuryFeb 9, 2023Break; Entrapment of Device

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2023, UTILIZING THE REFORM MC PEDICLE SCREW SYSTEM. WHILE ATTEMPTING TO ADJUST A SCREW, THE TIP OF THE ADJUSTABLE DRIVER (59-SP-0601) BROKE. THE BROKEN TIP PIECE WAS LEFT IN THE HEAD OF THE SCREW IMPLANTED IN THE PATIENT. THE PROCEDURE WAS COMPLETED UTILIZING OTHER INSTRUMENTATION, READILY AVAILABLE IN THE SET. THERE WAS NO DELAY TO THE PROCEDURE.

    Patient problems: Foreign Body In Patient

    MAUDE report 16336882 ↗