Skip to content

Veritas rTSA 3500-536U3Recalls & reported problems

Restor3d, Inc. · Humeral Liner, 5+, Ø36, +3, S · Class 2 · Implant

All models of Veritas rTSA (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840097513690), received between Jul 29, 2026 and Jul 29, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

1 other recall cover devices cleared under the same FDA submission (K243643); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Joint Dislocation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 29, 2026Migration

    A patient who had received a custom proximal humerus reconstruction reported experiencing a dislocation during physical therapy. The physician was not able to perform a closed reduction of the shoulder, so performed a surgery to correct it. It was discovered intraoperatively that the locking mechanism of the polymer liner had been damaged and had disassociated from the humeral implant. A new liner was implanted.

    Patient problems: Joint Dislocation

    MAUDE report 50072912 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K243643. One submission often covers many models, so check the product and lot details in each record.

  • Z-2987-2026 · Jun 30, 2026 · Open, Classified
    Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shel

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates