Skip to content

PlastiBell Circumcision Device 9211Recalls & reported problems

Aspen Surgical Products, Inc. · 1.1 cm PlastiBell Circumcision Device · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840113231546), received between Jul 16, 2025 and Sep 9, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Separation Failure1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    COMPLAINANT ALLEGES "DEVICE USED DURING CIRCUMCISION. PATIENT WAS DISCHARGED HOME. SUBSEQUENTLY PRESENTED TO THE EMERGENCY DEPARTMENT WITH POST-PROCEDURE BLEEDING. INITIAL EVALUATION NOTED A LACERATION TO THE GLANS PENIS DISTAL TO THE PLASTIBELL DEVICE; OBSERVATION WAS RECOMMENDED. LATER DURING HOSPITALIZATION, CONCERN AROSE THAT THE DEVICE HAD MIGRATED DISTALLY, POTENTIALLY COMPROMISING BLOOD FLOW. THE PATIENT WAS TAKEN TO THE OPERATING ROOM FOR DEVICE REMOVAL. INTRAOPERATIVELY, BLEEDING AND WORSENING HEMATOMA WERE NOTED. PATIENT WAS DISCHARGED HOME WITHOUT FURTHER KNOWN COMPLICATIONS."

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 23009006 ↗
  • InjuryJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    COMPLAINANT ALLEGES "CIRC PERFORMED ON 6/5. RETUNED WITHIN 12 HOURS OF THE CIRC WITH BLEEDING AND HEAD OF PENIS WAS SWOLLEN IN IT HAD TO BE SURGICALLY REMOVED."

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22521009 ↗
  • InjuryJul 16, 2025Separation Failure; Separation Failure

    COMPLAINANT ALLEGES "THE CIRCUMCISION WAS PERFORMED ON 6/6, INFANT DISCHARGED ON 6/7 WITH NORMAL ASSESSMENT POST CIRCUMCISION. INFANT RETURNED TO (B)(6) ED TODAY 6/17, DUE TO BLEEDING. UPON ASSESSMENT, THERE WAS AN ADHESION HOLDING THE PLASTIBELL ON, ONCE THE NNP WAS ABLE TO REMOVE THE ADHESION THE PLASTIBELL FELL OFF, NNP HELD PRESSURE AND BLEEDING RESOLVED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22520217 ↗