Skip to content

MedSource Catheters MS-86422Recalls & reported problems

Medsource International LLC · TrueSafe Comfort with Venapur, Blood Control 22G x 1” · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840155409750), received between Feb 24, 2025 and Feb 24, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 24, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    THE CATHETER CAME OFF THE PORT WHILE THE NURSE WAS REMOVING THE IV. THE PATIENT WAS TAKEN TO THE ER AND THEY HAD TO MAKE AN INCISION FOR REMOVAL UNDER ULTRASOUND.

    Patient problems: Foreign Body In Patient

    MAUDE report 21447341 ↗

Other Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days devices

All Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page