Skip to content

Blox™ SBT-321Recalls & reported problems

Boston Scientific Corporation · Endoscopic Bite Block with Elastic Cloth Strap · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840253108838), received between Mar 11, 2021 and May 13, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2026Material Separation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A BITE BLOX WAS USED IN THE MOUTH DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, A LOUD POPPING SOUND WAS HEARD. IT WAS DISCOVERED THAT THE PATIENT HAD BITTEN DOWN ON THE BITE BLOX, CAUSING A PIECE OF THE BITE BLOX TO BREAK OFF. THE PATIENT REQUIRED FURTHER EVALUATION USING A SECOND BRONCHOSCOPE TO CONFIRM THAT NO FRAGMENT HAD ENTERED THE AIRWAY OR LUNGS. THERE WERE NO PATIENT COMPLICATION AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25166441 ↗
  • MalfunctionMar 11, 2021Material Separation; Material Integrity Problem; Material Separation; Material Integrity Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A BITE BLOX WAS USED DURING AN UPPER ENDOSCOPY, INCLUDING GASTROSTOMY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTED THAT THE BITE BLOX WAS BROKEN INTO SEVERAL PIECES BY PATIENT'S FRONT TEETH AND THE HOSE OF THE GASTRIC CAMERA AS WELL AS THE SCOPE CHANNEL WERE FLATTENED. THE BROKEN PIECES WERE SUCCESSFULLY REMOVED FROM THE PATIENT'S MOUTH. THE PROCEDURE WAS COMPLETED WITH ANOTHER SCOPE AND ANOTHER OF THE SAME BITE BLOX. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11467002 ↗

Other Endoscopic bite block devices

All Endoscopic bite block devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.