This Boston Scientific Hedgehog double-end brush is used to clean the inside of endoscopes. Reports describe the brush tip or its bristles breaking off or detaching during cleaning, sometimes leaving pieces stuck inside the scope, which staff then retrieved with forceps or another brush. There are 32 reports, with no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00840253114020), received between Sep 18, 2018 and Sep 4, 2026 — classified as 34 malfunctions.
The most frequently reported specific device problem is “Break” (34 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break34
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- No Patient Involvement13
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50619134 ↗MalfunctionAug 21, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50374586 ↗MalfunctionApr 8, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2026. WHILE CLEANING THE ERCP SCOPE, THE BRUSH HEAD DETACHED AND ITS BRISTLES BECAME LODGED IN THE SCOPE. THE BRISTLES WERE SUCCESSFULLY RETRIEVED FROM THE SCOPE, AND THE CLEANING PROCESS WAS THEN COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24821494 ↗MalfunctionApr 9, 2025Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2025. DURING A SCOPE CLEANING, THE TIP OF THE BRUSH DETACHED AND REMAINED STUCK IN THE FORCEPS CHANNEL OF THE ENDOSCOPE. THE TIP WAS RETRIEVED USING ANOTHER CLEANING BRUSH AND FLUSHING WITH WATER. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21809113 ↗MalfunctionMar 12, 2025Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE BUNDLE WAS USED DURING A SCOPE CLEANING PERFORMED ON (B)(6) 2025. DURING THE SCOPE CLEANING, IT WAS NOTICED THAT THE BRISTLED PORTION OF THE BRUSH DETACHED INSIDE THE SUCTION VALVE UPON REMOVAL. THE DETACHED BRISTLE WAS RETRIEVED SUCCESSFULLY. THE SCOPE CLEANING WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INVOLVEMENT IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21584826 ↗MalfunctionAug 21, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2024. DURING SCOPE CLEANING, THE BRISTLES ON ONE SIDE OF THE DOUBLE END BRUSH BROKE. THE BRISTLES WERE NOT REMOVED FROM THE SCOPE. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG BRUSH. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20009168 ↗MalfunctionJul 23, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2024. DURING SCOPE CLEANING, THEY NOTICED THAT THE BRUSH DID NOT CAME OUT FROM THE TIP OF THE SCOPE. THEY REALIZED THAT THE BRUSH DETACHED INSIDE THE SCOPE. SEVERAL FLUSHES FROM ALL PORTS WERE DONE AND A TRADITIONAL METAL BRUSH WAS USED TO RETRIEVE THE BROKEN TIP. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG BRUSH. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19805817 ↗MalfunctionJun 25, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2024. DURING SCOPE CLEANING, THE BRUSH DETACHED INSIDE THE SCOPE. THE BRISTLE WAS REMOVED USING A GRASPING FORCEPS. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG BRUSH. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19610079 ↗MalfunctionJun 25, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2024. DURING SCOPE CLEANING, THE BRUSH DETACHED INSIDE THE SCOPE. THE BRISTLE WAS REMOVED USING A GRASPING FORCEPS. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG BRUSH. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19609965 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2024. DURING SCOPE CLEANING, THE BRUSH BRISTLE BROKE. THE BRISTLE WAS REMOVED USING FORCEPS. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG BRUSH. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18970311 ↗MalfunctionOct 2, 2023Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE BUNDLE WAS USED DURING A SCOPE CLEANING PERFORMED ON (B)(6) 2023. DURING THE CLEANING OF DUODENOSCOPE, THE BRUSH DETACHED FROM THE CLEAR PLASTIC HANDLE. THE DETACHED BRUSH WAS RETRIEVED BY HAND. THE SCOPE CLEANING WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INVOLVEMENT IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17851980 ↗MalfunctionSep 7, 2023Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2023. DURING THE SCOPE CLEANING, THE END OF THE CLEANING BRUSH BROKE OFF INSIDE THE ENDOSCOPE. THE DETACHED BRUSH WAS RETRIEVED BY PUSHING ANOTHER BRUSH THROUGH THE SCOPE. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17699788 ↗MalfunctionJan 4, 2023Break; Material Integrity Problem; Break; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2022. DURING SCOPE CLEANING THE END OF THE CLEANING BRUSH BROKE OFF INSIDE OF THE ENDOSCOPE. THE ENDOSCOPE WAS SENT FOR REPAIR IN ORDER TO RETRIEVE THE END OF THE BRUSH. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16094079 ↗MalfunctionOct 14, 2022Break; Material Integrity Problem; Break; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED ON AN UNKNOWN DATE. DURING SCOPE CLEANING, THE BRUSH BROKE AND WAS STUCK INSIDE THE ENDOSCOPE. THE ENDOSCOPE WAS SENT FOR REPAIR TO RETRIEVE THE BRUSH. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15607195 ↗MalfunctionMay 26, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING SCOPE CLEANING ON AN UNKNOWN DATE. DURING SCOPE CLEANING, THE STUBBY BRUSH OR SMALLER BRUSH BROKE. THE DETACHED BRUSH WAS RETRIEVED FROM THE SCOPE WITH HAND. THE PROCEDURE WAS COMPLETED WITH ANOTHER HEDGEHOG DOUBLE-END BRUSH. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14515266 ↗MalfunctionSep 9, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE BUNDLE WAS USED FOR SCOPE CLEANING ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING SCOPE CLEANING, THE BRUSH HEAD OF THE DOUBLE END CHANNEL BRUSH DISLODGED FROM THE HANDLE AND GOT STUCK IN THE BIOPSY CHANNEL. THE BRUSH HEAD WAS THEN REMOVED BY PUSHING IT OUT OF THE SCOPE USING BIOPSY FORCEPS. THE SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG DOUBLE BUNDLE. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12448865 ↗MalfunctionApr 20, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE BUNDLE WAS USED DURING SCOPE CLEANING ON AN UNKNOWN DATE. AFTER THE SCOPE AND CLEANING TRAY WENT THROUGH THE REPROCESSING MACHINE, A BRUSH BRISTLE WAS FOUND ON THE TRAY. THE BRISTLE WAS REMOVED BY WASHING THE TRAY. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11694446 ↗MalfunctionApr 20, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE-END BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6), 2021. ACCORDING TO THE COMPLAINANT, AFTER THE SCOPE WENT THROUGH THE REPROCESSING MACHINE, A BRISTLE WAS FOUND ON THE TIP OF THE SCOPE. THE BRISTLE WAS WIPED OFF WITH GAUZE. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11694441 ↗MalfunctionApr 20, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE BOUNDLE WAS USED DURING A SCOPE CLEANING PERFORMED ON AN UNKNOWN DATE. AFTER THE SCOPE WENT THROUGH THE REPROCESSING MACHINE, A BRUSH BRISTLE WAS FOUND ON THE NUMBER PLATE OF THE SCOPE. THE BRISTLE WAS REMOVED BY WIPING OFF WITH GAUZE. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11694436 ↗MalfunctionJan 13, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE SCOPE CLEANING, THE BRISTLES WERE STUCK ON THE BUTTON. THE TECHNICIAN REMOVED THE BRISTLES FROM THE BUTTON BY HAND. SCOPE CLEANING WAS COMPLETED USING ANOTHER HEDGEHOG BRUSH. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 11166022 ↗MalfunctionSep 17, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING SCOPE CLEANING, THE BRUSH WAS DETACHED AND A BRISTLE WAS LEFT INSIDE THE SCOPE. A PAIR OF FORCEPS WAS USED TO REMOVE THE BRISTLE FROM THE SCOPE. THE SCOPE CLEANING WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 10546807 ↗MalfunctionDec 9, 2019Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING SCOPE CLEANING, THE BRISTLES OF THE BRUSH DETACHED AND GOT STUCK IN TO THE SCOPE VALVES. THE BRISTLES WERE REPORTED TO HAVE BEEN REMOVED MANUALLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9437900 ↗MalfunctionNov 5, 2019Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DUAL-END BRUSH WAS USED DURING A SCOPE CLEANING ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, THE BRISTLES BROKE OFF IN THE SCOPE CHANNEL. ANOTHER BRUSH WAS USED TO COMPLETE THE CLEANING OF THE SCOPE. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED AS OF 07NOV2019. THE BRISTLES WERE REMOVED BY PUSHING IT OUT OF THE SCOPE CHANNEL.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 9278735 ↗MalfunctionSep 9, 2019Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DUAL-END BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, WHILE CLEANING THE SCOPE, THE CATHETER BROKE CAUSING THE BRUSH TO DETACH INSIDE THE WORKING CHANNEL OF THE SCOPE. THE BRUSH WAS REMOVED FROM THE SCOPE. IT IS UNKNOWN HOW THE SCOPE CLEANING WAS COMPLETED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8974194 ↗MalfunctionMay 22, 2019Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A HEDGEHOG DOUBLE BUNDLE BRUSH WAS USED DURING SCOPE CLEANING ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, WHILE CLEANING THE SCOPE, BRISTLES DETACHED FROM THE BRUSH HEAD AND GOT CAUGHT IN THE SPRING PART OF THE ASPIRATION BUTTON. IT WAS NOT REPORTED HOW THE DETACHED BRISTLES WERE REMOVED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8631821 ↗
Other ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE devices
- Hedgehog SBD-289Boston Scientific Corporation · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE129 reports
- Channel cleaning brush BW-20TOlympus Medical Systems Corp. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE90 reports
- MEDLINE DYK500FSAWMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- VERIFY LCC101Steris Corporation · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK500FSMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MetriSPONGE 10-4025Metrex Research, LLC · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK215FSMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK1000TSEMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page