EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-12.5/2.5) (REF VTICM5 12.6)Recalls & reported problems
Staar Surgical Company · VTICM5 - Diopter: -12.5; Cylinder 2.5; Overall Diameter (Length): 12.6 · Class 3 · Implant
All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00840311308712), received between Oct 15, 2023 and Aug 23, 2025 — classified as 10 injuries.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Inadequacy of Device Shape and/or Size2
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Inflammation1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT AN IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S EYE. THE PATIENT EXPERIENCED ASEPTIC INFLAMMATION, PIGMENT DEPOSITION. TREATMENT WAS PERFORMED. LENS REMAINS IMPLANTED. PATIENT IS STILL CURRENTLY BEING TREATED. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN. THIS REPORT IS PART OF A BOLUS SUBMISSION OF INDIVIDUALLY REPORTED EVENTS IDENTIFIED DURING A RETROSPECTIVE REVIEW BY THE CUSTOMER COVERING A SIX-MONTH PERIOD. EACH CASE IS SUBMITTED SEPARATELY AND DOES NOT INDICATE A TEMPORAL CLUSTER. THIS TIMELINE ALSO HIGHLIGHTS A DELAYED ONSET OF THE BOLUS REPORTED EVENTS RELATIVE TO THE PROCEDURE DATE (AVERAGE: 26 DAYS, RANGE:17-35 DAYS, AS WELL AS A DELAY IN REPORTING DUE TO RETROSPECTIVE CASE IDENTIFICATION. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Inflammation
MAUDE report 22880481 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED LOW VAULT. LENS REMAINS IMPLANTED. IN A SEPARATE A REPLACEMENT SHORTER LENGTH LENS WAS IMPLANTED AND PROBLEM RESOLVED. CAUSE OF THE EVENT IS REPORTED AS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22122711 ↗InjuryMar 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED A LENS ROTATION. ON SEPARATE VISIT THE LENS WAS REPOSITIONED, AND PROBLEM DID NOT RESOLVE. ON A THIRD VISIT THE LENS WAS EXCHANGED FOR A LONGER LENGTH LENS AND THE PROBLEM WAS RESOLVED. CAUSE REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21588134 ↗InjuryJan 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED A LOW VAULT WITH ROTATION. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR THE SAME MODEL, SHORTER LENGTH AND THE PROBLEM WAS RESOLVED. IT HAS BEEN NOTED THAT THE SURGEON SELECTED A DIFFERENT LENS SIZE THAN THAT INITIALLY SUGGESTED BY THE ICL CALCULATION SOFTWARE. THEREFORE, THERE IS NOT ENOUGH SAFETY AND EFFECTIVENESS DATA TO SUPPORT IMPLANTATION IN THIS PATIENT CATEGORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21180590 ↗InjuryNov 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED LOW VAULT WITHOUT ROTATION. IN A SEPARATE VISIT THE LENS WAS EXPLANTED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20768038 ↗InjurySep 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT AN IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S EYE. THE PATIENT EXPERIENCED LENS ROTATION NOT ASSOCIATED TO A LOW VAULT. THE LENS WAS EXCHANGED FOR A SAME LENGTH LENS AND THE PROBLEM RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20314595 ↗InjuryApr 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM VTICM5_12.6 IMPLANTABLE COLLAMER LENS OF -12.50/2.50/081 (SPHERE/CYLINDER/AXIS) WAS IMPLANTED INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6)2024. EXCESSIVE VAULT WAS OBSERVED. ON (B)(6) 2024 THE LENS WAS EXCHANGED FOR A SHORTER LENGTH LENS AND THE PROBLEM WAS RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19177108 ↗InjuryMar 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM, VTICM5_12.6, -12.5/2.5/100 (SPHERE/CYLINDER/AXIS), IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2024. LENS ROTATION NOT ASSOCIATED TO A LOW VAULT WAS OBSERVED, THE LENS WAS REPOSITIONED ON (B)(6) 2024, BUT DID NOT RESOLVE THE PROBLEM. ON 01/MAR/2024 THE LENS WAS REPLACED WITH A LONGER LENGTH LENS AND THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18979509 ↗InjuryOct 15, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VTICM5_12.6 IMPLANTABLE COLLAMER LENS OF DIOPTER -12.5/+2.5/092 INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2023. THE LENS WAS EXCHANGED ON (B)(6) 2023 FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT. THIS RESOLVED THE PROBLEM. THE REPORTER DID NOT INDICATE THE CAUSE OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17937006 ↗InjuryOct 15, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VTICM5_12.6 IMPLANTABLE COLLAMER LENS OF DIOPTER -12.5/+2.5/092 INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. THE LENS WAS EXCHANGED ON (B)(6) 2023 FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT. THIS RESOLVED THE PROBLEM. THE REPORTER DID NOT INDICATE THE CAUSE OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17936999 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
Other Phakic Toric Intraocular Lens devices
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 13.2 (-13.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-13.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-8.5/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.1 (-5.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-11.5/2.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-14.0/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-5.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-7.5/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.