Skip to content

EVO/EVO+VISIAN Implantable Collamer Lens VTICMO 13.2 (-15.0/2.0) (REF VTICMO 13.2)Recalls & reported problems

Staar Surgical Company · VTICMO - Diopter: -15.0; Cylinder 2.0; Overall Diameter (Length): 13.2 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840311327447), received between Aug 22, 2024 and Sep 11, 2026 — classified as 6 injuries.

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Intraocular Pressure Increased2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    A facility representative reported that an implantable collamer lens was implanted into the patient's eye. The patient experienced excessive vault and elevated IOP. The lens was later removed, and medication was provided. Cause of the event is reported as user error and other. There are multiple medical device reports associated with this patient. This report is 1 of 2 total reports. Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing-related issue or product deficiency.

    Patient problems: Intraocular Pressure Increased

    MAUDE report 50725291 ↗
  • InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem

    A facility representative reported that following an implantable collamer lens (ICL) implantation procedure, the patient experienced excessive vault, shallowing of AC and significant reduction of ICA. Lens exchanged with shorter length lens. Problem resolved. Cause of the event was reported as unknown.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50119386 ↗
  • InjuryJan 7, 2025No Apparent Adverse Event; No Apparent Adverse Event

    A HEALTHCARE PROFESSIONAL REPORTED FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED REFRACTIVE SURPRISE. LENS REMAINS IMPLANTED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCYADDITIONAL INFORMATION WAS REQUESTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS FILE WILL BE RE-OPENED IF ADDITIONAL INFORMATION IS PROVIDED.

    Patient problems: Intraocular Pressure Increased

    MAUDE report 21091021 ↗
  • InjuryOct 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UPON FURTHER REVIEW, IT HAS BEEN DETERMINED THAT THE EVENT IS NOT MDR REPORTABLE. THERE HAS BEEN NO REPORT OF SERIOUS INJURY OR MALFUNCTION. INITIAL MDR IS NOT APPLICABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20477030 ↗
  • InjurySep 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED LOW VAULT WITH ROTATION. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR A REPLACEMENT LENS OF LONGER LENGTH. THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20225672 ↗
  • InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT TWO WEEKS AFTER BILATERAL IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION INTO THE PATIENT'S EYE, INFLAMMATION OBSERVED BILATERALLY ON PO WEEK 2. INFORMATION REGARDING CONCOMITANT PRODUCTS USED INTRAOPERATIVELY WAS NOT PROVIDED. "DEPOSITS" (UNSPECIFIED) WERE NOTED ON THE ICL. NO FURTHER DESCRIPTION OF THE DEPOSITS WAS PROVIDED, AND DIAGNOSTIC CULTURES HAVE NOT BEEN REPORTED. IT WAS REPORTED THAT THE PATIENT EXPERIENCED NO OTHER REACTION AND THE LENS REMAINS IMPLANTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20027652 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.

Other Phakic Toric Intraocular Lens devices

All Phakic Toric Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page