Skip to content

FX V135 318-3212Recalls & reported problems

FX Shoulder Solutions, Inc. · FX V135 STEM TA6V Ø32/12 CEMENTLESS Ti/HA · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840452403642), received between May 4, 2026 and May 4, 2026 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Medical device site infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 4, 2026Adverse Event Without Identified Device or Use Problem

    SECOND REVISION OCCURRED ON (B)(6) 2026, APPROXIMATELY 10 MONTHS AFTER THE FIRST REVISION IN (B)(6) 2025. THE PRIMARY SURGERY HAPPENED ON (B)(6) 2025. NO FALL OR OTHER TRAUMA REPORTED. BASED ON COMPARING USAGE FORMS FX V135 STEM TA6V Ø32/12 CEMENTLESS TI/HA, HUMERAL CUP STANDARD PE/TA6V Ø36/+3, HUMELOCK REVERSED CENTERED GLENOSPHERE W/ SCREW COCR/TA6V 10° TILT Ø36MM, HUMELOCK REVERSED TA6V CEMENTLESS GLENOID BASEPLATE W/ SCREW TI/HA Ø24MM, +3MM LATERALIZED, CENTRAL SCREW TA6V Ø4.5MM L.20MM, STANDARD SCREW TA6V Ø4.5MM L.20MM, STANDARD SCREW TA6V Ø4.5MM L.35MM AND TWO LOCKING SCREW TA6V Ø4.5MM L.35MM WAS EXPLANTED IN MAY 2025 DUE TO INFECTION AND REPLACED WITH ANTIBIOTIC SPACER. THE ANTIBIOTIC SPACER WAS EXPLANTED, AND REPLACED WITH FX V135® HUMELOCK STEM TA6V Ø32/10 L. 120MM CEMENTLESS TI/HA, HUMERAL CUP STABILITY PE/TA6V Ø36/+3, HUMELOCK REVERSED CENTERED GLENOSPHERE W/ SCREW COCR/TA6V 10° TILT Ø36MM, HUMELOCK REVERSED TA6V CEMENTLESS GLENOID BASEPLATE W/PEG TI/HA Ø24MM, +3MM LATERALIZED, POST EXTENSION TA6V +06MM CEMENTLESS TI/HA, LOCKING SCREW TA6V Ø4.5MM L.15MM, LOCKING SCREW TA6V Ø4.5MM L.30MM, STANDARD SCREW TA6V Ø4.5MM L.30MM, CORTICAL SCREW TA6V Ø4.5MM L.28MM AND CORTICAL SC

    Patient problems: Medical device site infection

    MAUDE report 25070409 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page