Gram-Positive Blood Culture (BC-GP) Extraction Tray 20-009-018Recalls & reported problems

Luminex Corporation · The Verigene® Gram-Positive Blood Culture Nucleic Acid Test (BC-GP) performed using the sample-to-result Verigene® System is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840487101629), received between Oct 16, 2023 and Oct 16, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Volume Accuracy Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Volume Accuracy Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 16, 2023Volume Accuracy Problem; Volume Accuracy Problem

    (B)(4). UPON FURTHER INSPECTION, LOWER VOLUMES OF THE LABIASE ENZYME WERE OBSERVED IN WELL 2 OF THE UTILITY TRAY, AS SHOWN IN FIGURE 1 (ATTACHED). ONE OF THE TRAYS HAD A VISIBLY LOWER VOLUME OF REAGENT IN WELL 2, COMPARED TO THE PRODUCT SPECIFICATION OF 100±10 MICROLITERS. NC-04966 AND HRA-23-0033 WERE INITIATED FOR THE NONCONFORMING MATERIAL. THERE WERE NO FALSE RESULTS OBSERVED DURING FINAL PRODUCT TESTING FOR LOT 062123018C. THE LABIASE ENZYME IN THE VERIGENE BC-GP ASSAY IS AN ENZYME PREPARATION USEFUL FOR THE LYSIS OF ENTEROCOCCUS BACTERIA. WITHIN NC-04966, IT WAS DETERMINED THAT LOW VOLUMES OF THE LABIASE ENZYME COULD RESULT IN FALSE-NEGATIVE RESULTS, IMPACTING E. FAECALIS, E. FAECIUM, VANA, AND VANB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17940622 ↗