Skip to content

VascuTape Radiopaque Tape 1108-20Recalls & reported problems

Lemaitre Vascular, Inc. · Glow 'N Tell 55cm Tape (20 Strips) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840663100385), received between Mar 14, 2024 and Mar 14, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K932184); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Human-Device Interface Problem1

Patient problems

  • Pain1
  • Radiation Burn1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 14, 2024Human-Device Interface Problem; Human-Device Interface Problem

    TAPE WAS NOT REMOVED FROM PATIENT'S LEG FOR TWO WEEKS AFTER PROCEDURE. PATIENT UNDERWENT MRI WITH TAPE INADVERTANTLY ON HIS LEG, RESULTING IN PAIN AND REDNESS, AS WELL AS BURN OF THE TAPE MARKINGS AS AN IMPRINT ON THE PATIENT'S LEG AFTER TAPE WAS FINALLY REMOVED. PATIENT HAD ANGIO PROCEDURE IN IR SUITE ON (B)(6) 2024, FOLLOWED BY LYSIS CASE THE FOLLOWING DAY; ISSUE AROSE WHEN MRI CONDUCTED 2 WEEKS LATER ON (B)(6) 2024 WITH TAPE STILL ON. PATIENT FINALLY REMOVED TAPE FROM HIS LEG (B)(6) 2024, SEVERAL DAYS AFTER MRI, AND OBSERVED THE REDNESS AND MARKINGS. TAPE MEASUREMENT MARKINGS WERE BURNED INTO PATIENT'S LEG DURING MRI; EXCORIATED DOWN TOWARDS THE ANKLE, PEELING AND RED. PATIENT IS BEING TREATED WITH WOUND CARE AND ANTIBIOTICS AND WAS DISCHARTED BUT FOLLOWED UP WITH BY IR SUITE ON (B)(6) 2024.

    Patient problems: Pain; Radiation Burn; Skin Inflammation/ Irritation

    MAUDE report 18907870 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K932184. One submission often covers many models, so check the product and lot details in each record.

  • Z-1054-2013 · Mar 18, 2013 · Terminated
    LeMaitre Stent Guide: 1102-00 (100 strips); 1102-50 (50 strips); 1102-20 (20 strips). Intended to be placed on the skin to assist during imaging procedures.
  • Z-1055-2013 · Mar 18, 2013 · Terminated
    Glow 'N Tell 20 cm Tape: 1100-00 (100 strips); 1100-50 (50 strips); 1100-20 (20 strips). Intended to be placed on the skin to assist during imaging procedures.
  • Z-0434-2011 · Oct 22, 2010 · Terminated
    LeMaitre VascuTape Radiopaque Tape, LeMaitre Stent Guide, 100 strips Catalog Number: 1102-00 Intended to be placed on the patient's skin to assist during imaging procedures by ser
  • Z-0435-2011 · Oct 22, 2010 · Terminated
    LeMaitre VascuTape Radiopaque Tape, Glow 'N Tell 30cm Tape, 100 strips Catalog Number: 1100-00 Intended to be placed on the patient's skin to assist during imaging procedures by s
  • Z-0436-2011 · Oct 22, 2010 · Terminated
    LeMaitre VascuTape Radiopaque Tape, Glow 'N Tell 55cm Tape 20 strips Catalog Number: 1108-20 Intended to be placed on the patient's skin to assist during imaging procedures by ser
  • Z-0433-2011 · Oct 22, 2010 · Terminated
    LeMaitre VascuTape Radiopaque Tape, LeMaitre Stent Guide, 20 strips Catalog Number: 1102-20 Intended to be placed on the patient's skin to assist during imaging procedures by serv