Skip to content

Pruitt F3 Carotid Shunt 2011-10Recalls & reported problems

Lemaitre Vascular, Inc. · 10F Pruitt F3 Outlying Shunt with T-port · Class 2

All models of Pruitt F3 Carotid Shunt (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840663101191), received between Nov 26, 2019 and Mar 20, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K051067); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Inflation Problem1
  • Material Deformation1
  • Material Puncture/Hole1
  • Material Too Soft/Flexible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Blood Loss1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2026Material Deformation; Material Too Soft/Flexible

    IT WAS REPORTED THAT TWO PRUITT CAROTID SHUNTS EXPERIENCED ISSUES DURING INTERNAL CAROTID BALLOON INFLATION. IN THE FIRST CASE, THE INTERNAL CAROTID BALLOON RUPTURED ON THE FIRST INFLATION ATTEMPT. IN THE SECOND CASE, THE BALLOON DID NOT INFLATE UNIFORMLY AND APPEARED WEAK, LEADING THE PHYSICIAN TO DISCONTINUE USE OF THE DEVICE. A THIRD SHUNT WAS SUBSEQUENTLY USED AND FUNCTIONED AS INTENDED. NO PATIENT INJURY WAS REPORTED. ONE DEVICE WAS DISCARDED, AND IT IS UNCLEAR WHETHER THE SECOND DEVICE WILL BE RETURNED. NOTE: THIS REPORT IS 2 OF 2 RELATED TO THE SECOND SHUNT USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649823 ↗
  • MalfunctionMar 20, 2026Material Rupture

    IT WAS REPORTED THAT TWO PRUITT CAROTID SHUNTS EXPERIENCED ISSUES DURING INTERNAL CAROTID BALLOON INFLATION. IN THE FIRST CASE, THE INTERNAL CAROTID BALLOON RUPTURED ON THE FIRST INFLATION ATTEMPT. IN THE SECOND CASE, THE BALLOON DID NOT INFLATE UNIFORMLY AND APPEARED WEAK, LEADING THE PHYSICIAN TO DISCONTINUE USE OF THE DEVICE. A THIRD SHUNT WAS SUBSEQUENTLY USED AND FUNCTIONED AS INTENDED. NO PATIENT INJURY WAS REPORTED. ONE DEVICE WAS DISCARDED, AND IT IS UNCLEAR WHETHER THE SECOND DEVICE WILL BE RETURNED. NOTE: THIS REPORT IS 1 OF 2 RELATED TO THE FIRST SHUNT USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649806 ↗
  • MalfunctionFeb 9, 2023Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE BALLOON OF THE PRUITT F3 CAROTID SHUNT WOULD DEFLATE INTO THE SAFETY BALLOON. AS A RESULT, THE SHUNT WAS NOT TIGHT AND HAD TO BE REPLACED. NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16340463 ↗
  • MalfunctionFeb 19, 2021Material Rupture; Material Rupture

    THE INTERNAL CAROTID BALLOON OF THE F3 CAROTID SHUNT WAS DAMAGED DURING THE CAROTID ENDARTERECTOMY PROCEDURE. THE SURGEON THEN USED ANOTHER F3 CAROTID SHUNT AND CONTINUED THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11352820 ↗
  • InjuryNov 26, 2019Material Puncture/Hole; Material Puncture/Hole

    AT A BEGINNING OF A CAROTID ENDARTERECTOMY (CEA) PROCEDURE, USING A F3 CAROTID SHUNT, WHEN SURGEON ATTEMPTED TO OCCLUDE THE COMMON CAROTID ARTERY BY INFLATING THE COMMON CAROTID BALLOON WITH SALINE, HE OBSERVED THE BALLOON WAS LEAKING AND WAS UNABLE TO OCCLUDE THE COMMON CAROTID ARTERY RESULTING IN BLOOD LEAKAGE FROM THE ARTERY. SURGEON CLAMPED THE COMMON CAROTID ARTERY AND COMPLETED THE PROCEDURE WITHOUT ANY FURTHER COMPLICATIONS. SURGEON INSPECTED THE BALLOONS DURING PRE-USE CHECK AND FOUND THEM TO BE OPERATIONAL.

    Patient problems: Blood Loss

    MAUDE report 9377525 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051067. One submission often covers many models, so check the product and lot details in each record.

  • Z-0070-2012 · Oct 4, 2011 · Terminated
    LeMaitre 9F Pruitt F3 Outlying Carotid Shunt with T-port The Pruitt F3 Carotid Shunt (the Shunt) is a multi-lumen device with balloons at both the distal (internal carotid) and pr

Other Catheter, Intravascular Occluding, Temporary devices

All Catheter, Intravascular Occluding, Temporary devices →

Sources and updates