Pruitt F3 Carotid Shunt 2011-10Recalls & reported problems
Lemaitre Vascular, Inc. · 10F Pruitt F3 Outlying Shunt with T-port · Class 2
All models of Pruitt F3 Carotid Shunt (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840663101191), received between Nov 26, 2019 and Mar 20, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K051067); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture2
- Inflation Problem1
- Material Deformation1
- Material Puncture/Hole1
- Material Too Soft/Flexible1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Blood Loss1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Material Deformation; Material Too Soft/Flexible
IT WAS REPORTED THAT TWO PRUITT CAROTID SHUNTS EXPERIENCED ISSUES DURING INTERNAL CAROTID BALLOON INFLATION. IN THE FIRST CASE, THE INTERNAL CAROTID BALLOON RUPTURED ON THE FIRST INFLATION ATTEMPT. IN THE SECOND CASE, THE BALLOON DID NOT INFLATE UNIFORMLY AND APPEARED WEAK, LEADING THE PHYSICIAN TO DISCONTINUE USE OF THE DEVICE. A THIRD SHUNT WAS SUBSEQUENTLY USED AND FUNCTIONED AS INTENDED. NO PATIENT INJURY WAS REPORTED. ONE DEVICE WAS DISCARDED, AND IT IS UNCLEAR WHETHER THE SECOND DEVICE WILL BE RETURNED. NOTE: THIS REPORT IS 2 OF 2 RELATED TO THE SECOND SHUNT USED IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649823 ↗MalfunctionMar 20, 2026Material Rupture
IT WAS REPORTED THAT TWO PRUITT CAROTID SHUNTS EXPERIENCED ISSUES DURING INTERNAL CAROTID BALLOON INFLATION. IN THE FIRST CASE, THE INTERNAL CAROTID BALLOON RUPTURED ON THE FIRST INFLATION ATTEMPT. IN THE SECOND CASE, THE BALLOON DID NOT INFLATE UNIFORMLY AND APPEARED WEAK, LEADING THE PHYSICIAN TO DISCONTINUE USE OF THE DEVICE. A THIRD SHUNT WAS SUBSEQUENTLY USED AND FUNCTIONED AS INTENDED. NO PATIENT INJURY WAS REPORTED. ONE DEVICE WAS DISCARDED, AND IT IS UNCLEAR WHETHER THE SECOND DEVICE WILL BE RETURNED. NOTE: THIS REPORT IS 1 OF 2 RELATED TO THE FIRST SHUNT USED IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649806 ↗MalfunctionFeb 9, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE BALLOON OF THE PRUITT F3 CAROTID SHUNT WOULD DEFLATE INTO THE SAFETY BALLOON. AS A RESULT, THE SHUNT WAS NOT TIGHT AND HAD TO BE REPLACED. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16340463 ↗MalfunctionFeb 19, 2021Material Rupture; Material Rupture
THE INTERNAL CAROTID BALLOON OF THE F3 CAROTID SHUNT WAS DAMAGED DURING THE CAROTID ENDARTERECTOMY PROCEDURE. THE SURGEON THEN USED ANOTHER F3 CAROTID SHUNT AND CONTINUED THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11352820 ↗InjuryNov 26, 2019Material Puncture/Hole; Material Puncture/Hole
AT A BEGINNING OF A CAROTID ENDARTERECTOMY (CEA) PROCEDURE, USING A F3 CAROTID SHUNT, WHEN SURGEON ATTEMPTED TO OCCLUDE THE COMMON CAROTID ARTERY BY INFLATING THE COMMON CAROTID BALLOON WITH SALINE, HE OBSERVED THE BALLOON WAS LEAKING AND WAS UNABLE TO OCCLUDE THE COMMON CAROTID ARTERY RESULTING IN BLOOD LEAKAGE FROM THE ARTERY. SURGEON CLAMPED THE COMMON CAROTID ARTERY AND COMPLETED THE PROCEDURE WITHOUT ANY FURTHER COMPLICATIONS. SURGEON INSPECTED THE BALLOONS DURING PRE-USE CHECK AND FOUND THEM TO BE OPERATIONAL.
Patient problems: Blood Loss
MAUDE report 9377525 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051067. One submission often covers many models, so check the product and lot details in each record.
- Z-0070-2012 · Oct 4, 2011 · TerminatedLeMaitre 9F Pruitt F3 Outlying Carotid Shunt with T-port The Pruitt F3 Carotid Shunt (the Shunt) is a multi-lumen device with balloons at both the distal (internal carotid) and pr
Other Catheter, Intravascular Occluding, Temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page