Skip to content

Pruitt F3 Carotid Shunt 2012-12Recalls & reported problems

Lemaitre Vascular, Inc. · 9F Pruitt F3 Inlying Shunt with T-port · Class 2

All models of Pruitt F3 Carotid Shunt (10)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00840663101238), received between Sep 13, 2017 and Mar 30, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K051067); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem3
  • Leak/Splash3
  • Inflation Problem2
  • Material Rupture2
  • Mechanical Problem2
  • Degraded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient4

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Leak/Splash

    IT WAS REPORTED THAT DURING PRETESTING, THE COMMON CAROTID BALLOON (BLUE BALLOON) DID NOT INFLATE, SUGGESTING A POSSIBLE LEAK. NO ADDITIONAL COMPLICATIONS OR PATIENT INJURY WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24712793 ↗
  • MalfunctionJun 12, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PRUITT F3 CAROTID SHUNT BLUE BALLOON RUPTURED DURING TESTING FOR A CAROTID ENDARTERECTOMY PROCEDURE. THE DEVICE WAS NOT DIRECTLY USED ON THE PATIENT. NO INJURY WAS REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22197897 ↗
  • MalfunctionFeb 13, 2025Inflation Problem; Material Rupture; Inflation Problem; Material Rupture

    IT WAS REPORTED THAT THE PRUITT F3 SHUNT INTERNAL BALLOON DID NOT INFLATE DURING PRETESTING FOR A CAROTID ENDARTERECTOMY PROCEDURE. THE COMMON BLUE BALLOON WAS RIPPED. IT WAS NOT USED ON THE PATIENT. NO INJURY WAS REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21374732 ↗
  • MalfunctionDec 16, 2024Degraded; Mechanical Problem; Degraded; Mechanical Problem

    IT WAS REPORTED THAT THE PRUITT F3 SHUNT HAS A PUNCTURED PROXIMAL BALLOON. FOUND DURING PRE-USE TEST. NO INJURY REPORTED TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20941227 ↗
  • MalfunctionSep 24, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT WHEN THE PRUITT F3 WAS INSERTED AND INSTALLED, A HOLE WAS FOUND IN IT, MAKING IT IMPOSSIBLE TO USE. THE MATERIAL WAS UNIQUE, THEREFORE DOCTOR NEEDED TO CHANGE HIS TECHNIQUE TO COMPLETE THE PROCEDURE. DEVICE WAS NOT IN DIRECT CONTACT WITH THE PATIENT. NO INJURY WAS REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20294473 ↗
  • MalfunctionSep 18, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE PRUITT F3 SHUNT WHITE GUIDE WAS FOUND PUNCTURED DURING PRE-USE CHECK. IT WAS NOT USED ON THE PATIENT. NO INJURY WAS REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20249395 ↗
  • MalfunctionDec 22, 2021Inflation Problem; Inflation Problem

    ONE OF THE BALLOONS OF THE PRUITT F3 CAROTID SHUNT FAILED TO DEFLATE PROPERLY. USER DID NOT PROVIDE ANY ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13060052 ↗
  • MalfunctionJul 17, 2020Mechanical Problem; Mechanical Problem

    DURING PRE-USE CHECK, WHEN SURGEON ATTEMPTED TO INFLATE THE INTERNAL CAROTID BALLOON OF PRUITT F3 CAROTID SHUNT WITH SALINE, HE OBSERVED TWO HOLES IN THE BALLOON. THE SHUNT WAS THEN ISOLATED AND A NEW PRUITT F3 CAROTID SHUNT WAS USED FOR THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10291181 ↗
  • MalfunctionMay 2, 2019Deflation Problem; Deflation Problem

    AFTER SUCCESSFULLY COMPLETING THE CAROTID ENDARTECTOMY PROCEDURE, THE SURGEON ATTEMPTED TO DEFLATE THE INTERNAL CAROTID BALLOON BY ASPIRATING THE LIQUID INSIDE THE BALLOON USING A SYRINGE. HOWEVER, HE WAS UNABLE TO DEFLATE THE INTERNAL CAROTID BALLOON. AFTER ATTEMPTING MULTIPLE TIMES, HE WAS ABLE TO PARTIALLY DEFLATE THE BALLOON AND REMOVE IT FROM THE PATIENT'S CAROTID ARTERY. COMMON CAROTID BALLOON DEFLATED PROPERLY. THE SURGICAL TEAM HAD TESTED BOTH BALLOONS BEFORE THE PROCEDURE AND WERE FOUND TO BE FUNCTIONAL. NO FURTHER INTERVENTION WAS NEEDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8573478 ↗
  • MalfunctionApr 19, 2019Deflation Problem; Deflation Problem

    SURGEON WAS UNABLE TO REMOVE THE INTERNAL CAROTID BALLOON FROM THE PATIENT'S INTERNAL CAROTID ARTERY AFTER COMPLETION OF CAROTID ENDARTERECTOMY PROCEDURE. SO, HE PUNCTURED THE BALLOON WITH A NEEDLE TO DEFLATE THE BALLOON AND WAS THEN ABLE TO REMOVE IT. NO FURTHER INTERVENTION WAS NEEDED. THE SURGICAL TEAM TESTED THE DEVICE BEFORE THE PROCEDURE AND BOTH BALLOONS WERE FOUND TO BE FUNCTIONAL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8532733 ↗
  • MalfunctionSep 13, 2017Deflation Problem; Deflation Problem

    DURING A CAROTID ENDARTERECTOMY, THE SURGEON COULD NOT DEFLATE THE INTERNAL CAROTID BALLOON OF THE SHUNT. SURGEON THEN PULLED THE LUMEN OF THE SHUNT TO REMOVE THE INFLATED BALLOON.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6862598 ↗

FDA clearance and classification

Cleared through Special 510(k) K051067 on May 27, 2005 (decision: Substantially Equivalent). Submitted by LeMaitre Vascular, Inc. as “Pruitt F3 Outlying And Inlying Carotid Shunt (With And Without T-Port)”. The FDA review took 31 days. FDA record ↗

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K051067. One submission often covers many models, so check the product and lot details in each record.

  • Z-0070-2012 · Oct 4, 2011 · Terminated
    LeMaitre 9F Pruitt F3 Outlying Carotid Shunt with T-port The Pruitt F3 Carotid Shunt (the Shunt) is a multi-lumen device with balloons at both the distal (internal carotid) and pr

Common questions about the Pruitt F3 Carotid Shunt 2012-12

Has the Pruitt F3 Carotid Shunt 2012-12 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00840663101238), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the Pruitt F3 Carotid Shunt 2012-12?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Sep 13, 2017 and Mar 30, 2026. In those reports, the most common specific device problems are “Deflation Problem” (3), “Leak/Splash” (3) and “Inflation Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Pruitt F3 Carotid Shunt 2012-12?

None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Pruitt F3 Carotid Shunt 2012-12 FDA approved?

It was cleared through Special 510(k) K051067 on May 27, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Pruitt F3 Carotid Shunt 2012-12, and is it still sold?

The FDA's device database lists Lemaitre Vascular, Inc. as the company responsible for it (brand Pruitt F3 Carotid Shunt, model 2012-12). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Intravascular Occluding, Temporary devices

All Catheter, Intravascular Occluding, Temporary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page