Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00840663101290), received between Jun 4, 2019 and Sep 3, 2024 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Failure to Deflate” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K051067); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deflate4
  • Deflation Problem2
  • Fluid/Blood Leak2
  • Inflation Problem1
  • Unintended Deflation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient2
  • Hemorrhage/Blood Loss/Bleeding1
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjurySep 3, 2024Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE WHITE INTERNAL CAROTID BALLOON WOULD NOT INFLATE DURING THE PROCEDURE. A STENT WAS PLACED TO RESOLVE THE ISSUE. THE BLEEDING WAS ABLE TO BE CONTROLLED, AND NO HARM OCCURRED TO THE PATIENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20127683 ↗
  • MalfunctionMar 9, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED PRUITT F3 CAROTID SHUNT TUBE HAS HOLE IN IT. THE ISSUE WAS FOUND DURING PRE-USE CHECK AND NO INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16513554 ↗
  • MalfunctionNov 16, 2022Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THAT THE SURGEON HAD AN ISSUE WHILE USING A PRUITT F3 CAROTID SHUNT. AT THE END OF THE PROCEDURE, THE BALLOON WOULD NOT DEFLATE, SO THEY HAD TO MANUALLY POP THE BALLOON TO REMOVE THE DEVICE. THERE WAS NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15806115 ↗
  • MalfunctionSep 16, 2022Unintended Deflation; Unintended Deflation

    IT WAS REPORTED THAT THE PRUITT F3 CAROTID SHUNT WAS INFLATED IN THE INTERNAL ARTERY FOR A CEA PROCEDURE, BUT WHILE THE SURGEON WAS OPERATING THE BALLOON DEFLATED. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15439665 ↗
  • MalfunctionSep 16, 2022Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THAT THE SITE HAS AN ISSUE WITH THE PRUITT F3 CAROTID SHUNT WHERE THE BALLOON IS NOT DEFLATING AFTER REMOVING THE DEVICE FROM THE PACKAGE. THE ISSUE WAS NOTED PRIOR TO USE ON A PATIENT. NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15439634 ↗
  • MalfunctionApr 8, 2022Failure to Deflate; Failure to Deflate

    AFTER COMPLETING CAROTID ENDARTERECTOMY (CEA) PROCEDURE, WHEN THE SURGEON ATTEMPTED TO DEFLATE ONE OF THE BALLOONS OF THE PRUITT F3 CAROTID SHUNT, HE WAS UNABLE TO DO SO. SO, HE MANUALLY POPPED THE BALLOON TO REMOVE IT FROM THE PATIENT'S BLOOD VESSEL. THERE WAS NO HARM TO THE PATIENT BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14058977 ↗
  • MalfunctionApr 8, 2022Failure to Deflate; Failure to Deflate

    AFTER COMPLETING CAROTID ENDARTERECTOMY (CEA) PROCEDURE, WHEN THE SURGEON ATTEMPTED TO DEFLATE ONE OF THE BALLOONS OF THE PRUITT F3 CAROTID SHUNT, HE WAS UNABLE TO DO SO. SO, HE MANUALLY POPPED THE BALLOON TO REMOVE IT FROM THE PATIENT'S BLOOD VESSEL. THERE WAS NO HARM TO THE PATIENT BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14058792 ↗
  • MalfunctionMar 12, 2020Fluid/Blood Leak; Fluid/Blood Leak

    DURING PRE-USE CHECK, WHEN SURGEON ATTEMPTED TO INFLATE THE INTERNAL CAROTID BALLOON WITH SALINE, HE OBSERVED A LEAKAGE AT THE SAFETY BALLOON. DEVICE WAS NOT USED FOR THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9826199 ↗
  • MalfunctionJul 24, 2019Deflation Problem; Deflation Problem

    DURING PRE-USE CHECK, COMMON CAROTID BALLOON FAILED TO DEFLATE COMPLETELY.

    Patient problems: No Patient Involvement

    MAUDE report 8823219 ↗
  • MalfunctionJun 4, 2019Deflation Problem; Deflation Problem

    DURING THE CAROTID ENDARTERECTOMY PROCEDURE, THE SURGEON ATTEMPTED TO DEFLATE THE INTERNAL CAROTID BALLOON BY ASPIRATING THE LIQUID INSIDE THE BALLOON USING A SYRINGE. HOWEVER, HE WAS UNABLE TO DEFLATE THE INTERNAL CAROTID BALLOON. THE SURGICAL TEAM HAD TESTED BOTH BALLOONS BEFORE THE PROCEDURE AND BOTH WERE FOUND TO BE FUNCTIONAL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8667744 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051067. One submission often covers many models, so check the product and lot details in each record.

  • Z-0070-2012 · Oct 4, 2011 · Terminated
    LeMaitre 9F Pruitt F3 Outlying Carotid Shunt with T-port The Pruitt F3 Carotid Shunt (the Shunt) is a multi-lumen device with balloons at both the distal (internal carotid) and pr