Pruitt Aortic Occlusion Catheter 2100-12Recalls & reported problems
Lemaitre Vascular, Inc. · Pruitt Aortic Occlusion Catheter, 12F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840663101511), received between Aug 22, 2019 and Aug 5, 2022 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Perforation” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K872090); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Perforation2
- Fluid/Blood Leak1
- Leak/Splash1
- Material Integrity Problem1
- Material Puncture/Hole1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- No Patient Involvement2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2022Material Integrity Problem; Material Integrity Problem
WHEN THE PHYSICIAN OPENED THE PACKAGING, THE BALLOON HAD ALREADY RUPTURED IN THE TRAY. THE SURGEON THEN USED ANOTHER PURITT AORTIC OCCLUSION CATHETER TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15178197 ↗MalfunctionJun 10, 2020Material Puncture/Hole; Material Perforation; Material Puncture/Hole; Material Perforation
DURING PRE-USE CHECK, SURGEON NOTICED THE BALLOONS OF THE PRUITT AORTIC OCCLUSION CATHETERS TO BE POROUS. DEVICE WAS NOT USED FOR THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10138637 ↗MalfunctionMar 20, 2020Mechanical Problem; Mechanical Problem
PRUITT AORTIC OCCLUSION CATHETER WAS USED FOR OCCLUDING AORTA DURING ABDOMINAL AORTIC ANEURYSM REPAIR (AAA). SOON AFTER OCCLUDING THE ARTERY WITH THE INFLATED BALLOON OF THE OCCLUSION CATHETER, SURGEON OBSERVED SALINE LEAKAGE FROM THE WHITE STOPCOCK DEFLATING THE BALLOON. THERE WAS NO INJURY TO THE PATIENT AS THE RESULT OF THE DEVICE MALFUNCTION SURGEON USED ANOTHER PRUITT AORTIC OCCLUSION CATHETER AND CONTINUED THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9859553 ↗MalfunctionMar 16, 2020Leak/Splash; Leak/Splash
DURING PRE-USE CHECK, SURGEON OBSERVED A LEAKAGE AT THE TIP OF THE BALLOON WHEN HE ATTEMPTED TO INFLATE THE BALLOON WITH SALINE. DEVICE WAS NOT USED FOR THE PROCEDURE.
Patient problems: No Patient Involvement
MAUDE report 9837275 ↗MalfunctionSep 27, 2019Material Perforation; Material Perforation
DURING AN OPEN STENT-GRAFT PLACEMENT FOR TREATING ANEURYSM OF A DESCENDING AORTA, AFTER PROPERLY PLACING THE STENT GRAFT IN HIS AORTA, SURGEON INSERTED AN AORTIC OCCLUSION CATHETER INTO THE STENT GRAFT. HE THEN INFLATED THE BALLOON WITH SALINE AT HIS DESIRED LOCATION TO OCCLUDE THE VESSEL BUT IT IMMEDIATELY RUPTURED. THERE WAS NO INJURY TO THE PATIENT AS THE RESULT OF THIS INCIDENT. SURGEON REPLACED THE AORTIC OCCLUSION CATHETER AND CONTINUED THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9130893 ↗MalfunctionAug 22, 2019Fluid/Blood Leak; Fluid/Blood Leak
DURING PRE-USE CHECK, SURGEON DETECTED A LEAKAGE AT THE JOINT WHERE THE INFLATION ARM MEETS THE IRRIGATION LUMEN. DEVICE WAS NOT USED IN THE PATIENT. SURGEON COMPLETED THE PROCEDURE USING ANOTHER PRUITT AORTIC OCCLUSION CATHETER THAT THEY HAD IN STOCK.
Patient problems: No Patient Involvement
MAUDE report 8920512 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K872090. One submission often covers many models, so check the product and lot details in each record.
- Z-0219-06 · May 17, 2005 · TerminatedProduct is a Pruitt Aortic Occlusion Catheter contained in a molded tray. The product is labeled in part: ''***LeMaitre VASCULAR***Distributed By:LeMaitre Vascular, Inc.***Manufa
Other Occluder, Catheter Tip devices
- Pruitt Irrigation Occlusion Catheter 2102-09Lemaitre Vascular, Inc. · Occluder, Catheter Tip1 recall4 reports
- Pruitt Irrigation Occlusion Catheter e2102-09Lemaitre Vascular, Inc. · Occluder, Catheter Tip1 recall0 reports
- Polyurethane Dual Lumen Occlusion Catheter 5F-2L-40Lemaitre Vascular, Inc. · Occluder, Catheter Tip1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page