Skip to content

Pruitt Occlusion Catheter 2103-56Recalls & reported problems

Lemaitre Vascular, Inc. · Pruitt Occlusion Catheter, 5F · Class 2

All models of Pruitt Occlusion Catheter (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840663101573), received between Mar 8, 2024 and Mar 8, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K811276); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 11, 2025Class IIOpen, Classified

    During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

    Z-1752-2025 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 8, 2024Material Rupture; Material Rupture

    THE PRUITT OCCLUSION CATHETER (POC) AND XENOSURE WERE USED IN ONE CASE OF ENDARTERECTOMY + EVT. AFTER ENDARTERECTOMY OF THE RIGHT FEMORAL ARTERY AND PLACEMENT OF A METAL STENT, THE SHAFT OF THE POC WAS INSERTED INTO THE VESSEL TO PLACE A PATCH AT THE ENDARTERECTOMY SITE. HOWEVER, THE BALLOON OF THE POC RUPTURED, SO THE BLOOD FLOW COULD NOT BE BLOCKED. SINCE THE STENT WAS PLACED ON THE PROXIMAL SIDE OF THE ENDARTERECTOMY SITE, IT WAS NOT POSSIBLE TO OCCLUDE THE BLOOD VESSEL USING VASCULAR CLAMP. SO, PTA BALLOON WAS USED INSTEAD TO OCCLUDE THE BLOOD FLOW. THE XENOSURE PATCH WAS SUTURED TO THE ENDARTERECTOMY SITE AND THE SURGERY WAS COMPLETED. THE PATIENT IS DOING WELL AND THERE WERE NO HEALTH COMPLICATIONS. THE CAUSE OF THE BALLOON RUPTURE APPEARS TO BE CONTACT WITH THE METAL PART OF THE STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18865689 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K811276. One submission often covers many models, so check the product and lot details in each record.

  • Z-1751-2025 · Apr 11, 2025 · Open, Classified
    Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56

Other Catheter, Flow Directed devices

All Catheter, Flow Directed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page