Pruitt Occlusion Catheter 2103-56Recalls & reported problems
Lemaitre Vascular, Inc. · Pruitt Occlusion Catheter, 5F · Class 2
All models of Pruitt Occlusion Catheter (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840663101573), received between Mar 8, 2024 and Mar 8, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K811276); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 11, 2025Class IIOpen, Classified
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
Z-1752-2025 · Root cause: Packaging process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 8, 2024Material Rupture; Material Rupture
THE PRUITT OCCLUSION CATHETER (POC) AND XENOSURE WERE USED IN ONE CASE OF ENDARTERECTOMY + EVT. AFTER ENDARTERECTOMY OF THE RIGHT FEMORAL ARTERY AND PLACEMENT OF A METAL STENT, THE SHAFT OF THE POC WAS INSERTED INTO THE VESSEL TO PLACE A PATCH AT THE ENDARTERECTOMY SITE. HOWEVER, THE BALLOON OF THE POC RUPTURED, SO THE BLOOD FLOW COULD NOT BE BLOCKED. SINCE THE STENT WAS PLACED ON THE PROXIMAL SIDE OF THE ENDARTERECTOMY SITE, IT WAS NOT POSSIBLE TO OCCLUDE THE BLOOD VESSEL USING VASCULAR CLAMP. SO, PTA BALLOON WAS USED INSTEAD TO OCCLUDE THE BLOOD FLOW. THE XENOSURE PATCH WAS SUTURED TO THE ENDARTERECTOMY SITE AND THE SURGERY WAS COMPLETED. THE PATIENT IS DOING WELL AND THERE WERE NO HEALTH COMPLICATIONS. THE CAUSE OF THE BALLOON RUPTURE APPEARS TO BE CONTACT WITH THE METAL PART OF THE STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18865689 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K811276. One submission often covers many models, so check the product and lot details in each record.
- Z-1751-2025 · Apr 11, 2025 · Open, ClassifiedPruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56
Other Catheter, Flow Directed devices
- Graftmaster 1012580-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED30 deaths225 reports
- SWAN-GANZ CCOMBO V 777F8Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls92 reports
- Graftmaster 1012581-16Abbott Vascular Inc. · CATHETER, FLOW DIRECTED21 deaths147 reports
- SWAN-GANZ CCOMBO V 774F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls2 deaths41 reports
- Graftmaster 1012581-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED12 deaths135 reports
- Graftmaster 1012582-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths77 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall23 reports
- Graftmaster 1012580-26Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths72 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page