TEC 6 PLUS VAP DES 120V ESRecalls & reported problems
Datex-Ohmeda Inc · Class 2
All models of TEC 6 PLUS (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datex-Ohmeda Tec 6 Plus vaporizer is used to deliver anesthetic gas during surgery. The 63 reports mostly describe units found during the recall inspection to deliver less anesthetic agent than the dial setting showed, with no patient involved. There are 63 malfunctions, no deaths and no injuries. One open recall, started November 2024, covers this reduced-concentration problem. Reports dropped sharply, from 62 in the prior 12 months to 1 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 63
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 00840682112789), received between Jan 15, 2025 and Nov 13, 2025 — classified as 63 malfunctions.
The most frequently reported specific device problem is “Output below Specifications” (63 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K000275); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 18, 2024Class IIOpen, Classified
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Z-0814-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Output below Specifications63
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions63
Recent adverse event reports
The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23543497 ↗MalfunctionOct 31, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434104 ↗MalfunctionOct 28, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23406472 ↗MalfunctionOct 28, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23406444 ↗MalfunctionOct 22, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23360687 ↗MalfunctionOct 22, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23360384 ↗MalfunctionOct 22, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23357083 ↗MalfunctionOct 18, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23332068 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325507 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325486 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325485 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325471 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325397 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325198 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325151 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325141 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325096 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23325064 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23324950 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23322572 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23322571 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23322568 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23322130 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23321162 ↗MalfunctionOct 17, 2025Output below Specifications; Output below Specifications
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN AGENT DELIVERY LESS THAN 20% FROM THE SETTING DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23320764 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K000275. One submission often covers many models, so check the product and lot details in each record.
- Z-2161-2013 · Jul 29, 2013 · TerminatedTec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthe
- Z-0201-2007 · Sep 15, 2006 · TerminatedDatex-Ohmeda Tec 6 Plus Desflurane Vaporizer. It is an electronic vaporizer which delivers the anesthetic agent desflurane.
Other VAPORIZER, ANESTHESIA, NON-HEATED devices
- TEC 850 SEVDatex-Ohmeda Inc · VAPORIZER, ANESTHESIA, NON-HEATED1 recall2,251 reports
- TEC 850 ISODatex-Ohmeda Inc · VAPORIZER, ANESTHESIA, NON-HEATED1 recall184 reports
- TEC 820 SEVDatex-Ohmeda Inc · VAPORIZER, ANESTHESIA, NON-HEATED1 recall59 reports
- TEC 6 PLUS VAP DES 220/240V ENDatex-Ohmeda Inc · VAPORIZER, ANESTHESIA, NON-HEATED1 recall48 reports
- TEC 820 ISODatex-Ohmeda Inc · VAPORIZER, ANESTHESIA, NON-HEATED1 recall31 reports
- Vapor 2000 M35054Drägerwerk AG & Co. KGaA · Vaporizer, Anesthesia, Non-Heated1 recall8 reports
- Vapor 3000 M36500Drägerwerk AG & Co. KGaA · Vaporizer, Anesthesia, Non-Heated1 recall1 report
- TEC 6 PLUS VAP DES 120V USADatex-Ohmeda Inc · VAPORIZER, ANESTHESIA, NON-HEATED36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page