1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K180374); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jan 31, 2025Open, Classified

    For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

    Z-1239-2025 · Root cause: Software design · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K180374. One submission often covers many models, so check the product and lot details in each record.

  • Z-1965-2019 · Jul 2, 2019 · Terminated
    Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D

Other Transducer, ultrasonic, diagnostic devices

All Transducer, ultrasonic, diagnostic devices →