Skip to content

Vivid iq v203Recalls & reported problems

GE Medical Systems (China) Co.,Ltd. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682139694), received between Feb 21, 2025 and Feb 21, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fire” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fire1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 21, 2025Fire; Fire

    IT WAS REPORTED THAT AFTER COMPLETING AN ULTRASOUND-GUIDED PROCEDURE, THE ULTRASOUND DEVICE WAS MOVED ASIDE. THE DEVICE REMAINED CONNECTED TO THE MAIN POWER SUPPLY OUTLET AND WAS NOT IN USE AT THE TIME OF THE EVENT. CUSTOMER REPORTED AFTER ABOUT 54 MINUTES, FLAMES WERE NOTICED TO BE ORIGINATING FROM THE POWER SUPPLY LOCATED ON THE SYSTEM. ONCE THE POWER CORD WAS UNPLUGGED FROM THE MAIN SUPPLY OUTLET, THE FLAME EXTINGUISHED ITSELF WITHIN 20 SECONDS. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21439404 ↗

Other System, imaging, pulsed echo, ultrasonic devices

All System, imaging, pulsed echo, ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page