NaturaLyte 13-3251-9Recalls & reported problems
Fresenius Medical Care Renal Therapies Group, LLC · NaturaLyte Liquid Acid Concentrate for Bicarbonate Dialysis, Drum · Class 2
All models of NaturaLyte (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840861101887), received between Nov 30, 2020 and Dec 2, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Chemical Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 14, 2026Class IIOpen, Classified
Complaints of blue-green discoloration in product
Z-3092-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Chemical Problem2
Patient problems
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2020Chemical Problem; Chemical Problem
A HEMODIALYSIS (HD) CLINIC BIOMEDICAL TECHNICIAN REPORTED TO FRESENIUS CUSTOMER SERVICE THAT THEY HAD A NATURALYTE LOT THAT RESULTED IN LOW CONDUCTIVITY VALUES ON AN UNSPECIFIED NUMBER OF HD MACHINES. UPON FOLLOW UP THE REGISTERED NURSE (RN) CONFIRMED THE REPORTED EVENT. SHE CLARIFIED THAT THE OCCURRENCE DATES ARE UNKNOWN. SHE CONFIRMED THAT IN ALL CASES WHERE THEY EXPERIENCED CONDUCTIVITY ISSUES DURING TREATMENT, ALL THE PATIENTS WERE ABLE TO COMPLETE THEIR TREATMENT AFTER SWITCHING THE JUGS OF NATURALYTE MID-TREATMENT. THERE WERE NO ADVERSE EVENTS, SERIOUS INJURIES, NOR REQUIRED MEDICAL INTERVENTION AS A RESULT OF THESE EVENTS. THEY ARE CURRENTLY USING BIBAGS BI-CARB. SAMPLE RETURN PICKUP WAS SCHEDULED. THIS RECORD CAPTURES ALL OF THE CLINIC¿S UNSPECIFIED NUMBER OF INCIDENTS OF LOW CONDUCTIVITY EXPERIENCED DURING TREATMENT WHILE UTILIZING THE REPORTED BATCH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10933143 ↗MalfunctionNov 30, 2020Chemical Problem; Chemical Problem
A HEMODIALYSIS (HD) CLINIC BIOMEDICAL TECHNICIAN REPORTED TO FRESENIUS CUSTOMER SERVICE THAT THEY HAD TWO NATURALYTE BATCHES THAT HAD LOW CONDUCTIVITY VALUES DURING SETUP PRIOR TO HD TREATMENT. UPON FOLLOW UP THE REGISTERED NURSE (RN) CONFIRMED THE REPORTED EVENT. SHE INDICATED THEY HAVE CONTINUED TO USE THE AFFECTED BATCHES AFTER RECEIVING AN GUIDANCE ON PRESCRIPTION, AND SETTING CHANGES THAT WILL AVOID THE CONDUCTIVITY ISSUES. THEY ARE CURRENTLY USING BIBAGS BI-CARB. SAMPLE RETURN PICKUP WAS SCHEDULED. THEY EXPERIENCED LOW CONDUCTIVITY DURING SETUP AN UNSPECIFIED NUMBER OF TIMES AND ON UNKNOWN DATES. THIS RECORD CAPTURES ALL OF THE CLINIC¿S INCIDENTS OF LOW CONDUCTIVITY EXPERIENCED DURING SETUP WHILE UTILIZING THE REPORTED BATCH.
Patient problems: No Patient Involvement
MAUDE report 10919444 ↗
Other DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER) devices
- NaturaLyte 08-3251-9Fresenius Medical Care Renal Therapies Group, LLC · DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)2 recalls78 reports
- NaturaLyte 08-2251-0Fresenius Medical Care Renal Therapies Group, LLC · DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)3 recalls1 death20 reports
- NaturaLyte 08-2301-3Fresenius Medical Care Renal Therapies Group, LLC · DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)2 recalls35 reports
- Dry Acid Dissolution Unit 160147Fresenius Usa, Inc. · DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)129 reports
- GranuFlo 0FD2251-3BFresenius Medical Care Renal Therapies Group, LLC · DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)2 recalls1 death20 reports
- GranuFlo 0FD3251-3BFresenius Medical Care Renal Therapies Group, LLC · DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)2 recalls1 death1 report
- Acidified Concentrate Distribution and Storage (ACDS) System MD704262Isopure, Corp. · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)2 recalls0 reports
- Bicarby RFP-400-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record