TIGER SHARK SYSTEM S-PT50-A2402Recalls & reported problems

Choice Spine, LP · TIGERSHARK M,STERILE,SPACER,24LX2H · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840996188791), received between Aug 26, 2025 and Aug 26, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    DR. IMPLANTED TSM DEVICES IN THE L5/S1 AND L4/L5 DISC SPACES ON 4/21/25. HE USED 24 MM LENGTH, 6DEGREE ENDPLATES (S-PT50-E2406; LOT 11861-03). HE USED A 2MM SPACER (S-PT50-A2402; LOT 10455-19) IN THE L4/5 DISC SPACE AND A 3MM SPACER IN THE L5/S1 DISC SPACE. POSTOPERATIVE IMAGES TAKEN AT THE END OF THE CASE SHOWED MISALIGNMENT OF THE INFERIOR TSM ENDPLATE AND THE TSM SPACER. AT A LATER DATE, DR. DECIDED TO REVISE THE HARDWARE AND COMPLETED A REVISION SURGERY FOR THIS PATIENT ON (B)(6) 2025. THE TSM CAGE WAS REMOVED AND REPLACED WITH A DIFFERENT INTERBODY. THE REMOVED IMPLANTS WERE DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22898146 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page