N/A 09-0341Recalls & reported problems
Biomet Microfixation, Inc · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841036028862), received between Sep 24, 2022 and Sep 24, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2022Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT FRACTURED DURING A PROCEDURE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. ANOTHER INSTRUMENT WAS USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15475591 ↗
FDA clearance and classification
The FDA regulates product code EMJ (elevator, surgical, dental) under 21 CFR 872.4565 as a Class I device. 21 CFR 872.4565 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the N/A 09-0341
Has the N/A 09-0341 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00841036028862), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the N/A 09-0341?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 24, 2022. That report lists device problem “Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A 09-0341?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A 09-0341 FDA approved?
Devices of this type (product code EMJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Biomet Microfixation, Inc.
Who makes the N/A 09-0341, and is it still sold?
The FDA's device database lists Biomet Microfixation, Inc as the company responsible for it (brand N/A, model 09-0341). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELEVATOR, SURGICAL, DENTAL devices
- N/A 09-0257Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL57 reports
- N/A 09-0252Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL49 reports
- N/A SP-2359Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL28 reports
- N/A 09-0313Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL17 reports
- N/A 09-0262Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL15 reports
- N/A 09-0386Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL9 reports
- N/A 09-0258Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL8 reports
- N/A 09-0259Biomet Microfixation, Inc · ELEVATOR, SURGICAL, DENTAL8 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page