HTR-PMI PM622996Recalls & reported problems
Biomet Microfixation, Inc · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841036282981), received between Apr 3, 2020 and Apr 3, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).
1 other recall cover devices cleared under the same FDA submission (K924935); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THE CUSTOM CRANIAL PLATE WAS TOO LARGE AND HAD TO BE ALTERED INTRAOPERATIVELY PRIOR TO IMPLANTATION. THE CUSTOM PLATE WAS SUCCESSFULLY IMPLANTED ALTHOUGH THE SURGEON DOES NOT BELIEVE THE RESULT IS PERFECT DUE TO THE IMPLANT BEING TOO THICK, ESPECIALLY IN THE FRONTAL AREA. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9920598 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K924935. One submission often covers many models, so check the product and lot details in each record.
- Z-2751-2016 · Jun 13, 2016 · TerminatedHard Tissue Replacement (HTR) Implant, Rx only, for use in single procedure only. Custom made. HTR polymer implants are intended for bone replacement/augmentation for treatment of
Other Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction devices
- OmniPore Surgical Implant OP7544Matrix Surgical Holdings, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction1 recall1 report
- OmniPore Surgical Implant OP9520Matrix Surgical Holdings, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction1 recall1 report
- OmniPore Surgical Implant OP9519Matrix Surgical Holdings, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction1 recall1 report
- OmniPore Surgical Implant OP7543Matrix Surgical Holdings, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction1 recall0 reports
- Su-Por Surgical Implants 4353Poriferous, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction8 reports
- OmniPore Surgical Implant OP8330Matrix Surgical Holdings, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction3 reports
- HTR-PEKK PK620905Oxford Performance Materials, Inc. · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction2 reports
- OmniPore Surgical Implant OP8322Matrix Surgical Holdings, LLC · Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page