Response Spine 5.5/6.0 System 00-1003-4001Recalls & reported problems
Orthopediatrics Corp. · LRG. SET SCREWS · Class 2 · Implant
All models of Response Spine 5.5/6.0 System (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00841132100103), received between May 1, 2019 and Jun 2, 2026 — classified as 3 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Migration” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration3
- Detachment of Device or Device Component2
- Device Dislodged or Dislocated1
- Fracture1
- Loose or Intermittent Connection1
- Mechanical Problem1
Patient problems
- Failure of Implant5
- No Information2
- Insufficient Information1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 2, 2026Migration
PATIENT PRESENTED IN CLINIC 1 YEAR POST OP AND FOUND MULTIPLE SET SCREWS DISENGAGED.
Patient problems: Failure of Implant
MAUDE report 25364722 ↗MalfunctionMay 21, 2026Migration
A POST-OPERATIVE DEVICE-RELATED ISSUE WAS IDENTIFIED IN A PATIENT WHO UNDERWENT SPINAL SURGERY ON (B)(6) 2024, UTILIZING A RESPONSE 5.5/6.0 SCOLIOSIS SYSTEM. THE IMPLANTED DEVICE INCLUDED A LARGE SET SCREW (PART NUMBER: 00-1003-4001). ON (B)(6) 2026, DURING A ROUTINE CLINIC FOLLOW-UP VISIT, THE PATIENT PRESENTED WITH A DISENGAGED SET SCREW. THE EVENT WAS DISCOVERED IN A POST-OPERATIVE CLINICAL SETTING. THERE WAS NO REPORTED DEATH OR PATIENT INJURY ASSOCIATED WITH THIS EVENT; HOWEVER, THE DISENGAGEMENT RESULTED IN THE FAILURE OF THE ORIGINAL SURGICAL CONSTRUCT. AS A RESULT OF THIS DEVICE ISSUE, A REVISION SURGERY HAS BEEN DEEMED NECESSARY, AND SCHEDULING IS PENDING. AT THIS TIME THE DEVICE REMAINS IMPLANTED IN THE PATIENT, AND THEREFORE, NO PRODUCT IS CURRENTLY AVAILABLE FOR RETURN OR EVALUATION.
Patient problems: Failure of Implant
MAUDE report 25241560 ↗MalfunctionMar 24, 2026Migration
PATIENT HAD ORIGINAL SURGERY ON (B)(6) 2025. IN POST OP FILMS BACK IN (B)(6), NO ISSUES SEEN. FOLLOW-UP VISIT WHEN KID SHOWED UP AT CLINIC, AT LEAST 5 SET SCREWS DISENGAGED ON RIGHT SIDE. 6.0.
Patient problems: Failure of Implant
MAUDE report 24681790 ↗MalfunctionJul 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
KYPHOSIS CASE DONE AROUND 4 YEARS AGO. PATIENT PRESENTS WITH BROKEN ROD. DURING REVISION, SURGEONS TRY TO EXTEND 1 LEVEL BILATERAL BUT SURGEON BLOWS PEDICLE ON R SO ONLY EXTEND ON L BUT REPLACE RODS. DURING REDUCTION, A LOT OF LORDOSIS IN ROD. PATIENT COMES BACK LESS THAN WEEK LATER WITH SET SCREW DISENGAGED ON THE EXTENDED LEVEL. ADD SIDE BY SIDE TO R SIDE AND REPLACE DISENGAGED SET SCREW.
Patient problems: Failure of Implant
MAUDE report 22490284 ↗MalfunctionJun 19, 2025Loose or Intermittent Connection; Loose or Intermittent Connection
ON (B)(6), DR. (B)(6) STARTED TO IMPLANT RESPONSE 6.0 INTO A PATIENT. DUE TO PATIENT HEALTH ISSUES THE CASE WAS ABORTED WITH ABOUT HALF THE SCREWS IMPLANTED. HE THEN CAME BACK ON (B)(6) AND COMPLETED THE SURGERY WHICH INCLUDED LEVELS T2 TO L2. ON (B)(6), DR. (B)(6) REVISED THE PATIENT DUE TO THE LOWER SET SCREWS HAVING FALLEN OUT. AFTER EXPOSING HE FOUND THAT NINE OF THE LOWER 10 SET SCREWS WERE LOOSE OR MISSING. HE WAS VERY CONCERNED ABOUT THE SCREWS LOOSENING FARTHER PROXIMAL IN THE CONSTRUCT. HE SAID HE UNDERSTANDS THE BOTTOM TO LOOSENING, BUT DID NOT UNDERSTAND WHY THEY LOOSENED THE REST OF THE WAY UP. HE WOULD LIKE TO TALK TO SOMEBODY ABOUT TESTING WHETHER BURNING OR CUTTING THE ROD INSITU HAS THE POTENTIAL TO LOOSEN THE ROD.
Patient problems: Failure of Implant
MAUDE report 22290319 ↗MalfunctionFeb 7, 2025Mechanical Problem; Mechanical Problem
SURGEON NOTICED THAT SET SCREWS HAD BACK OUTED AND HAS SUBSEQUENTLY REVISED . THIS WAS SURGEONS FIRST CASE EITHER RESPONSE, AND DID NOT USE THE REDUCTION TOWER AT THE BOTTOM OF THE CONSTRUCT. HANDLES TORQUE WERE RETURNED AND FOUND TO BE IN CALIBRATION.
Patient problems: Insufficient Information
MAUDE report 21333255 ↗InjuryFeb 21, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT HAS BEEN REPORTED THAT APPROXIMATELY THREE YEARS FOLLOWING THE PLACEMENT OF A RESPONSE SPINAL CONSTRUCT, THE PATIENT UNDERWENT A REVISION DUE TO ROD FRACTURE AND LOOSE SET SCREWS. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Information
MAUDE report 9738473 ↗InjuryFeb 21, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT HAS BEEN REPORTED THAT APPROXIMATELY THREE YEARS FOLLOWING THE PLACEMENT OF A RESPONSE SPINAL CONSTRUCT, THE PATIENT UNDERWENT A REVISION DUE TO ROD FRACTURE AND LOOSE SET SCREWS. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Information
MAUDE report 9738456 ↗InjuryMay 1, 2019Fracture; Fracture
IT WAS REPORTED THAT FOLLOWING A POSTERIOR FUSION SURGERY, THE PATIENT EXPERIENCED BACK PAIN AND CLICKING. IT WAS DISCOVERED DURING THE REVISION PROCEDURE THAT A SET SCREW WAS LOOSE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Pain
MAUDE report 8571189 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page