Skip to content

eZ-Fix 800-200Recalls & reported problems

Orthopediatrics Corp. · Rail, Mono 200mm Blue · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132142837), received between Oct 15, 2024 and Oct 15, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    DATE OF INITIAL SURGERY: (B)(6) 2024. DETAILS: RIGHT ULNA ACUTE CORRECTION, EXCISION OF OSTEOCHONDROMA, AND APPLICATION OF RAIL FIXATOR. SURGEON: (B)(6). COMPLICATION: LOOSENING OF PROXIMAL CLAMP OF RAIL CAUSING LOSS OF CORRECTION 6 WEEKS INTO LENGTHENING PROCESS. THIS WAS PICKED UP ON X-RAY ONE WEEK PRIOR TO THE REVISION SURGERY AND THE PATIENT HAD NOTICED THE CLAMP HAD SHIFTED AND WAS NO LONGER PERPENDICULAR TO THE HALF PINS. WHEN WE REVISED THE RAIL FIXATOR INTRA-OP WE NOTED THAT ALL OF THE LOCKING MECHANISMS ON THE PROXIMAL CLAMP WERE LOOSE SO SUSPECT FAILURE OF THIS CLAMP ALTOGETHER.

    Patient problems: Failure of Implant

    MAUDE report 20458153 ↗

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page