Skip to content

Cabo™ 81-0234Recalls & reported problems

Neurostructures, Inc · 2 LEVEL PLATE 34mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841508101376), received between Jan 26, 2018 and Jan 26, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Mechanical Problem1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2018Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    SURGEON HAD TO REVISE AN ACDF PATIENT THAT WAS JUST OVER 3 MONTHS OUT. PHYSICIAN NOTED AT 3 WEEKS THAT ONE OF THE SCREWS ON THE BOTTOM OF THE PLATE HAD BACKED OUT "ONE TURN". HE DID NOT SURGICALLY INTERVENE AT THAT TIME. THE PATIENT LATER REPORTED HEARING A POP (APPROXIMATELY 13 WEEKS POST-OP). THE PATIENT WAS SEEN BY THE SURGEON AFTER THIS EVENT. THE X-RAY REVEALED THE BOTTOM SCREW HAD BACKED OUT FURTHER, AND BOTH OF THE TOP SCREWS WERE BROKEN CLOSE TO MID-SHAFT. THE SURGEON TOOK THE PATIENT BACK TO SURGERY AND REMOVED THE PLATE AND SCREWS, EXCEPT FOR THE 2 BROKEN TIPS WHICH HE LEFT IN THE BONE. THE SURGEON CHOSE NOT TO RE-PLATE. THE PATIENT WAS PUT IN A HARD COLLAR AND WILL BE MONITORED TO SEE IF FURTHER INTERVENTION IS NEEDED.

    Patient problems: Failure of Implant

    MAUDE report 7222713 ↗

Other Appliance, Fixation, Spinal Intervertebral Body devices

All Appliance, Fixation, Spinal Intervertebral Body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page