Cabo™ 81-1214Recalls & reported problems
Neurostructures, Inc · Variable - 4.0mm Self Tapping 14mm · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841508101666), received between Oct 27, 2016 and Jan 26, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient2
- Failure of Implant1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2018Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
SURGEON HAD TO REVISE AN ACDF PATIENT THAT WAS JUST OVER 3 MONTHS OUT. PHYSICIAN NOTED AT 3 WEEKS THAT ONE OF THE SCREWS ON THE BOTTOM OF THE PLATE HAD BACKED OUT "ONE TURN". HE DID NOT SURGICALLY INTERVENE AT THAT TIME. THE PATIENT LATER REPORTED HEARING A POP (APPROXIMATELY 13 WEEKS POST-OP). THE PATIENT WAS SEEN BY THE SURGEON AFTER THIS EVENT. THE X-RAY REVEALED THE BOTTOM SCREW HAD BACKED OUT FURTHER, AND BOTH OF THE TOP SCREWS WERE BROKEN CLOSE TO MID-SHAFT. THE SURGEON TOOK THE PATIENT BACK TO SURGERY AND REMOVED THE PLATE AND SCREWS, EXCEPT FOR THE 2 BROKEN TIPS WHICH HE LEFT IN THE BONE. THE SURGEON CHOSE NOT TO RE-PLATE. THE PATIENT WAS PUT IN A HARD COLLAR AND WILL BE MONITORED TO SEE IF FURTHER INTERVENTION IS NEEDED.
Patient problems: Failure of Implant
MAUDE report 7222745 ↗MalfunctionDec 20, 2016Detachment of Device or Device Component; Detachment of Device or Device Component
INITIAL ACDF SURGERY AT C4-C6 WAS PERFORMED ON (B)(6) 2016. ROUTINE FOLLOW UP AT 2 WEEKS NOTED NO CHANGE TO THE DEVICE. APPROXIMATELY 6 WEEKS POST OPERATIVELY, IT WAS NOTED THAT THE 18MM CAUDAL BONE SCREW HAD BACKED OUT 2-3MM FROM THE ANTERIOR CERVICAL PLATE. PATIENT IS ASYMPTOMATIC. SURGEON IS MONITORING THE PATIENT, AND NO REVISION SURGERY IS CURRENTLY PLANNED. PATIENT BONE QUALITY IS UNKNOWN. THE PATIENT REPORTEDLY DID NOT FALL PRIOR TO THE EVENT. IT IS UNKNOWN IF THE PATIENT COMPLIED W/POST-OPERATIVE CARE INSTRUCTIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6190072 ↗MalfunctionOct 27, 2016Detachment of Device or Device Component; Detachment of Device or Device Component
INITIAL ACDF SURGERY AT C4-C6 WAS PERFORMED ON (B)(6) 2016. ROUTINE FOLLOW UP AT 2 WEEKS NOTED NO CHANGE TO THE DEVICE. APPROXIMATELY 6 WEEKS POST OPERATIVELY IT WAS NOTED THAT THE 18 MM CAUDAL BONE SCREW HAD BACKED OUT 2-3MM FROM THE ANTERIOR CERVICAL PLATE. PATIENT IS ASYMPTOMATIC. SURGEON IS MONITORING THE PATIENT, AND NO REVISION SURGERY IS CURRENTLY PLANNED. PATIENT BONE QUALITY IS UNKNOWN. THE PATIENT REPORTEDLY DID NOT FALL PRIOR TO THE EVENT. IT IS UNKNOWN IF THE PATIENT COMPLIED W/POST-OPERATIVE CARE INSTRUCTIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6062584 ↗
Other Appliance, Fixation, Spinal Intervertebral Body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page