Collamer Ultraviolet Absorbing Posterior Chamber Intraocular Lens CQ2015A (UDI 00841542101882)Recalls & reported problems
Staar Surgical Company · Diopter: 15.0; Aspheric; Three Piece · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841542101882), received between May 17, 2017 and Dec 12, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Abrasion From Instrument Or Another Object” (1 report).
5 other recalls cover devices cleared under the same FDA submission (P990013); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Abrasion From Instrument Or Another Object1
- Device Operates Differently Than Expected1
- Improper or Incorrect Procedure or Method1
- No Apparent Adverse Event1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2018No Apparent Adverse Event; No Apparent Adverse Event
THE REPORTER INDICATED THE SURGEON INSERTED A CQ2015A COLLAMER THREE PIECE LENS, +15.0 DIOPTER, AND THE LENS HAPTIC TORE. THE LENS WAS REMOVED DURING THE SAME SURGERY, WITH NO PATIENT INJURY. THE BACKUP LENS WAS IMPLANTED AND THE PROBLEM WAS RESOLVED. THE REPORTER INDICATED THE CAUSE OF THE EVENT WAS UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8158092 ↗MalfunctionAug 30, 2017Improper or Incorrect Procedure or Method; Device Operates Differently Than Expected; Improper or Incorrect Procedure or Method; Device Operates Differently Than Expected
THE REPORTER INDICATED THAT AS THE SURGEON WAS ATTEMPTING TO LOAD A 13.0MM CQ2015A IMPLANTABLE COLLAMER LENS DIOPTER +15.0, THE SURGEON HAD A PROBLEM LOADING THE LENS. THE REPORTER STATED NO PATIENT INJURY. UPON LENS RETURN, THE BOX HAD THE WORDS "PATIENT CONTACT" WRITTEN ON IT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN UNSUCCESSFUL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6832599 ↗MalfunctionMay 17, 2017Device Abrasion From Instrument Or Another Object; Device Abrasion From Instrument Or Another Object
THE REPORTER STATED THAT A CQ2015A, +15.0 DIOPTER, THREE PIECE COLLAMER LENS WAS IMPLANTED AND EXPLANTED DURING THE SAME SURGERY DUE TO A NICK IN THE OPTIC. THE NICK OCCURRED DURING THE LOADING PROCESS DUE TO THE USER NOT USING THE OPTIMAL TYPE OF FORCEPS FOR THIS TYPE OF LENS. THE REPORTER STATED THIS WAS THE FIRST TIME THE USER WAS PERFORMING A PROCEDURE FOR THIS TYPE OF LENS. THERE WAS NO ENLARGEMENT OF THE INCISION AND NO SUTURES NEEDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6574310 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990013. One submission often covers many models, so check the product and lot details in each record.
- Z-1747-2013 · Jun 10, 2013 · TerminatedAfinity CQ2015A Intraocular Lens The visual correction of aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction.
- Z-1746-2013 · Jun 10, 2013 · TerminatednanoFLEX CC4204A Intraocular Lens Correct aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction. The IOL is to be implante
- Z-2706-2011 · May 20, 2011 · TerminatedBrand Name STAAR Surgical Collamer Ultraviolet-Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens Common Names Collamer single piece IOL Collamer plate haptic IOL
- Z-1037-04 · Jun 17, 2004 · TerminatedSTAAR Surgical Intra Occular Lens, Collamer.
- Z-0539-04 · Feb 9, 2004 · TerminatedCollamer UV One-Piece IntraOcular Lens. Models 4205BF and 4204BF
Other Intraocular Lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page